Date
Re: first post - new home lab - tips on using HP/Agilent gear
Thanks for the precise instructions - I will do that this coming weekend after another session of home renovations in the new place is complete. I'm bookmarking this.
By Samudra N3RDX S21X · #126563 ·
Re: first post - new home lab - tips on using HP/Agilent gear
Thank you Michael, yes, I think they are 75-ohm, so ¡­ at short distances (in the same rack) what can I do to match the 50-ohm interfaces with the test equipment. I bought this years ago, and
By Samudra N3RDX S21X · #126562 ·
Re: first post - new home lab - tips on using HP/Agilent gear
Oh, forgot about the need to terminate the _unused_ outputs, been a long time. I grew up with thick ethernet, CSMA/CD 10 Mbps, so should have remembered the water pipe flow paradigm. So, if we
By Samudra N3RDX S21X · #126561 ·
Re: 53181A 3 GHz prescaler from China
Thanks for the input.? I ordered one from China.? I will see what happens.?? As Lothar said, it is worth $40 to take a chance.? I am sure they were all built off the same plans available on
By Tom B · #126560 ·
Re: HP 8662A Power up - overvoltage shutdown
My guess would be that something else in the control loop is faulty. Because it kind sounds like it is working ok in a steady state condition,?but not to transients. Or is conditionally
By EI2KU · #126559 ·
Re: HP 8662A Power up - overvoltage shutdown
Hi Tony I should have said. the mains voltage sensors are VR1 and VR2.? Use ONLY a none grounded instrument such as a hand held DMM to make mesurements around them.? Use of Grounded instruments
By george edmonds · #126558 ·
Re: HP 8662A Power up - overvoltage shutdown
Hi Tony The overvoltage sensors are VR1 and VR2, has one gone faulty? G Edmonds
By george edmonds · #126557 ·
Re: HP 8662A Power up - overvoltage shutdown
A question or so: 1) is the equipment supposed to run at 230 VAC?? Are you in a region where you could get 240?? Is it specified for that? 2) any idea if it's a supply triggering this, or a
By Harvey White · #126556 ·
Re: HP 8662A Power up - overvoltage shutdown
Digging a bit more I find that the PSU will start successfully at 220V AC and at 225V AC. But at 230V AC and above it throws and over-voltage error. Not quite sure where to take it now... Tony
By Tony · #126555 ·
Re: first post - new home lab - tips on using HP/Agilent gear
If I'm not mistaken, the designer of the 8671/8672 is (or at least was) on this mailing list. Perhaps he'll chime in. -Dave
By Dave McGuire · #126554 ·
Re: HP 8662A Power up - overvoltage shutdown
I have not measured it to be frank. But both of the above tests were with the variac set to 230V - either powering on the instrument and turning it up gradually, or otherwise setting the variac to
By Tony · #126553 ·
Re: HP 8662A Power up - overvoltage shutdown
Hi Tony What is your actual mains voltage? G Edmonds
By george edmonds · #126552 ·
HP 8662A Power up - overvoltage shutdown
Hi, I'm hoping somebody can advise me how to approach a power-on issue with my HP 8662A signal generator. It came with a faulty power supply and I went through a detailed fault finding process and
By Tony · #126551 ·
Re: 53181A 3 GHz prescaler from China
Many years ago I remade/copied the original Agilent card. It works fine. I think the Chinese did the same thing industrially.
By maurit · #126550 ·
Re: first post - new home lab - tips on using HP/Agilent gear
Are the video amplifiers 50 or 75 ohm?
By Michael A. Terrell · #126549 ·
Re: Frequency range of HP 5352B frequency counter
Hi Dave, There is some info in these posts: /g/HP-Agilent-Keysight-equipment/message/111396 /g/HP-Agilent-Keysight-equipment/message/103664 Seems as the only
By Askild · #126548 ·
Re: HP 3465A offset when input shorted
On Mon, May 23, 2022 at 04:23 PM, Chris Smith wrote: > > Hi all, > > Just obtained a rather sick HP3465A (the 4.5 digit one, not the big 6.5 > digit unit). It is functioning but there is a one
By Ozan · #126547 ·
Re: first post - new home lab - tips on using HP/Agilent gear
Generally use the distribution amplifier outputs it¡¯s the best way to do it and also make sure to terminate the unused outputs ! As far as the 8671B concerns, turn on the unit and let it warm up,
By Lothar baier <Lothar@...> · #126546 ·
Re: 53181A 3 GHz prescaler from China
I bought one a while back for my 53132A. I haven't done much testing of it, but it seems to work just fine. Keith
By Keith · #126545 ·
Re: HP 8640 Bad Frequency Counter
Sounds like one or more of the fingers in the range switch have let go of their supporting disc. If that's the case, they're probably wandering around in the bottom of the cabinet so carefully
By n4buq · #126544 ·