Re: My $25 "In Poor Shape" 3468A has arrived
Meant to say on DCV the readings with the jumper fitted are between 0.000000 and 0.000001
By Alex · #95859 ·
Re: E4401B No 50MHZ Cal signal
Hello Alex, Everything is automated during alignment so no settings will affect this. Sounds like the unit has an issue and yes similar to discussions here. It's easy to check - with a cable from 50
By Mark Bielman · #95858 ·
Re: My $25 "In Poor Shape" 3468A has arrived
So my new nice zero ohm jumpers arrived, and here are the results. Just plugged in the jumper into the input terminals and on power up the DCV indicates between 0.000001 and 0.000001. When I go to 2W
By Alex · #95857 ·
Re: E4401B No 50MHZ Cal signal
Sorry to barge in with this, but I have my sights on a 4411B (I assume it is similar to the 4401B being discussed here) that has a message on the startup screen that says "Align RF Skipped, No align
By Alex · #95856 ·
Re: HP 8644B Option 2 Upgrade
Well, thought Colby had the answer but when I looked and the I/O board does not have U32, I had to look at the service manual again. This is only for 8664/65 models. So stuck again. Mark
By Mark Bielman · #95855 ·
Re: HP8560E Low Level
Ok, but if the generator goes to 3GHz, you might just be within the high band. My 8562A has high band from 2.75GHz (low band is up to 2.9GHz). It has coax relay between high and low, so it doesn't
By Askild · #95854 ·
Re: HP8560E Low Level
Askild Unfortunately the signal generator I have goes to 3GHz and also the Spectrum Analyser also only goes to 3GHz (found an Anritsu Spectrum Master I forgot was a SA as well). I did inject 310.7MHz
By Dave Ireland · #95853 ·
Re: hp8640B_with_rough_sound
Yes, the problem is solved. The Problem was a short circuit inside the main oscillator to the Varactor lines. Initially I checked all supplies >> No Problem. Then the voltage that goes to the
By Armin Roesch · #95852 ·
Re: HP 8644B Option 2 Upgrade
I know that unit shares a lot with 8664/8665 units, and if the attenuator is the same, your attenuator cal ROM only has cal data up to ~1GHz. If memory serves me right, it's U32 on the backplane/IO
By Colby Burkett · #95851 ·
Re: 141T varying CRT intensity?
Can you set it up to measure the bias voltage (cathode to control grid) WHILE OPERATING - be good to find out what is changing.
By Bruce · #95850 ·
HP 8644B Option 2 Upgrade
Hello group. I have an 8644B (standard, upto 1030 MHz) that I wanted to upgrade to 2060 MHz. Managed to locate an upgrade kit w/cables (thanks Steve!) but after installing it I get the following
By Mark Bielman · #95849 ·
Re: HP 85044A T/R test set part - Cabinet Trim "zipper" - 08502-20007 - source or 3D printed copy?
Dave, Is it possible for you to share a photo of the case where the zipper locks fit? The service manual on the Keysight site is scanned from a poor microfiche. ¡ªGlen
By Glen Hoag · #95848 ·
Re: 141T varying CRT intensity?
...And it occurs to me that the Brightness pot would be the first thing to look at, since it has 'moving parts', a wiper, etc. If the pot is proven to be OK, then use it as a starting point, and
By saipan59 (Pete) · #95847 ·
Re: 141T varying CRT intensity?
I suggest a cold/cracked solder joint could be the issue - influenced by flexing/shifting when heat builds up. If it "jumps around" between good and bad, that might rule out a component that has
By saipan59 (Pete) · #95846 ·
Re: hp3586a selective voltmeter jammed
Yes, my copy is poor as well.? I suppose you can check the startup circuit near U5 to start. Maybe Artek has a decent scan. Peter
By Peter Gottlieb · #95845 ·
Re: hp3586a selective voltmeter jammed
Yes, I just fixed the power supply some days ago and the three voltages (+5, +12V, -12V) sounds good even though I didn't check the ripple. I have the service manual the microprocessor parts
By cc@... · #95844 ·
Re: HP 85044A T/R test set part - Cabinet Trim "zipper" - 08502-20007 - source or 3D printed copy?
Hello, just send what you need by registered mail to my office in Budapest. But: A) my oven can not handle large pieces B) I am semi-retired. So things can take time, as I have so many jobs... ---
By Tam Hanna · #95843 ·
Re: hp3586a selective voltmeter jammed
It looks like a digital control problem.? Check the power supplies for voltage and ripple first.? Do you have the service manual?
By Peter Gottlieb · #95842 ·
Re: HP 85044A T/R test set part - Cabinet Trim "zipper" - 08502-20007 - source or 3D printed copy?
I can try that. Right now, I'm working on making a board positioner for the binocular microscope that reads positions of parts from the board file. I don't have any cases that need work until I start
By Harvey White · #95841 ·
hp3586a selective voltmeter jammed
Hello, ? have you ever experienced this fault? This unit is now stuck at this leve. Earlier it was jammed as well but only with annunciator LEDs (no LED display) then I moved and disconnected the
By cc@... · #95840 ·