¿ªÔÆÌåÓý

Date
Re: 7470A Plotter
You raise an interesting question. I once borrowed a manual on the HP 5370B time interval counter from Poul-Henning Kamp, since I wanted one and at the time there was not one on the Agilent web site.
By David Kirkby <david.kirkby@...> · #49081 ·
Re: designspark
This not free - about $75, but it is very good. http://www.abacom-online.de/uk/html/sprint-layout.html Great email support Ken
By ken_holladay <jacquiken@...> · #49080 ·
Re: 8563E +28V shutdown/oscillates
IF it takes seconds to cycle on and off, suspect that it's overheating and protecting itself. The junction temp would be in the 150-160C range, so the IC would be really, really hot. From your last
By dnmeeks · #49079 ·
Re: designspark
Hi Jerry, For work, I use Altium, Viewdraw (now Mentor PADS) and OrCad. Each has its own features & quirks. For play I use Express PCB (find it great for even just documenting small hand-wired
By John Dusatko · #49078 ·
Re: 7470A Plotter
Better get it quick the copy they have posted is an "illegal" ArtekMedia Scan :-( I thought those guys has better taste than that Dave ArtekManuals -- Dave
By Artekmedia <manuals@...> · #49077 ·
Re: 7470A Plotter
At the Australian HPcomputer museum www.hpmuseum.net. -Rik
By Rik Bos · #49076 ·
Re: designspark
I use DIPTRACE, a free version is available which is quite powerful. I decided on this package on recommendation of a colleague who has much experience and has tried almost everything
By rovmkr · #49075 ·
7470A Plotter
Does any one have a copy of the Users Manual for this Plotter? I have found the Interface Manual on line but not the User Manual. I may get one to use with my "new" 8559A just for a test/fun. I even
By Chuck · #49074 ·
Re: designspark
Hi Jerry, The DesignSpark support is only via email and is really only for software issues. Direct help is mostly in the user group forum. http://www.designspark.com/discuss/index.php Try looking
By Susan Parker · #49073 ·
Re: designspark
I use express PCXB. The learning curve is very short. Easy to make custom symbols. 73 Glenn WB4UIV
By Glenn Little · #49071 ·
Problem with HP Scopes 54820A Power Supply Unit - please help
Hi, I have a problem with my HP Scopes 54820A Power Supply Unit and I need a schematic ofr it. Can anybody help me with it. HP Part no 0950-3252 (IPS Inc PN NX300-6003) Thank you in advance
By micro_debugger <elcom@...> · #49072 ·
Re: designspark
Don't use it. Eagle, for small stuff, hobby license for large stuff, non-profit is guaranteed. Some like it, some hate it. Harvey
By Harvey White · #49070 ·
Re: Failing 3585A Spectrum Analyzer
Hi Gordon, When you change the span, other things in the analyzer change too. The resolution bandwidth and video bandwidth probably change at the 500kHz span setting as well. That means you can do
By pianovt · #49069 ·
Re: HP 8648A Error code 517
Hi Cslim, Thanks, I sure hope you get your generator working correctly. Regards, Jeff ________________________________ From: Chin SiangLim <cslim388@...> To:
By Jeff Miller <jmelect@...> · #49068 ·
Re: HP 8648A Error code 517
How true. ________________________________ To: hp_agilent_equipment@... Sent: Sat, January 12, 2013 11:47:30 AM Subject: Re: [hp_agilent_equipment] HP 8648A Error code 517 That's an
By Jeff Miller <jmelect@...> · #49067 ·
Re: HP 8648A Error code 517
Hi jeff, That is just great! Happy for you. I will check mine. Will also let you know when I am able to get the schematics. Cslim
By cslim388 · #49066 ·
Re: Failing 3585A Spectrum Analyzer
Many SAs that I know of change operating modes somewhere in their span range, from the LO being fully and continuously synthesized, to using open-loop linear sweep tuning. Once the span is large
By Ed Breya · #49065 ·
Re: HP 8648A Error code 517
That's an old trick. Radios used to present for service with tubes swapped around all the time. Well, they fit don't they. Don Black.
By Jaanette Black · #49064 ·
Re: HP 8648A Error code 517
Hi Cslim, You wouldn't believe what I FOUND. I thought I would check for the obvious. What I found was 2 of the modules were interchanged. They are the RF OUT and the SIG GEN SYNTH. I purchased
By Jeff Miller <jmelect@...> · #49063 ·
Re: Failing 3585A Spectrum Analyzer
Hi Gordon, This is almost certainly related to the way the LO is structured. Generally speaking, the LO has to make some changes as the span increases. In your case, for spans that are 500 kHz and
By pianovt · #49062 ·