CHECK OUT THE WIKI The purpose of the wiki is mainly to allow you to find information on instruments, either from either
- The model number
- The function(s) listed below. Some instruments have multiple functions - for example, the 4195A is a VNA, spectrum analyzer and an impedance analyzer. Therefore the 4195A is listed in multiple categories
Please also check out HPWiki available here:
- Accessory kits - various types
- AC power analyzers - PA2201A and PA2203A
- AC power supplies 6811C, 6812C, 6813C
- Airlines
- Arbitrary waveform generators M8194A
- Amplifiers?493A, 495A?
- Attenuators (optical) 8156A, 8157A, 8158B, 81566A, 81576A,?
- Attenuators (RF) 8494A
- Attenuator set (500 ¦¸) 350C
- Attenuator set (600 ¦¸) 350D
- Attenuator switch driver
- Audio analyzers? 8903A, 8903B, 8903E,? ?
- Base station test sets
- Bit error rate testers (BERTs)
- Cables
- Capacitance meters U1701A, U1701B, 4272A, 4278A, 4279A
- Capacitor Bridge 4270A,
- Capacitor standards 16380A, 16380C,?
- Carrier noise test setsi
- Cesium frequency standards
- Clamp ammeters
- Close field probes
- Crystal Impedance E4915A, E4916A
- Data Acquisition Systems (DAQs)
- DC power analyzers
- DC power supplies 6030A , 6031A , 6032A, 6033A, 6035A, 6131C, 6621A, 6622A, 6623A, 6624A, 6627A, 6255A, 6645A, 6671A, 6672A, 6673A, 6674A, 6675A, 62003A, 62003C, 62003E, 62004A, 62004B, 62004E, 62005A, 62005B, 62005E, 62006A, 62006B, 62006E, 62010A, 62010C, 62010E, 62012A, 62012C, 62012E, 62015A, 62015C, 62015E, 62018A, 62018C, 62018E, 62024A, 62024C, 62024E, 62028A, 62028C, 62028E, 62048A, 62048C, 62048E
- Delay lines
- Detectors
- Device current waveform analyzers
- Digital communications analyzers
- Directional couplers
- Distortion analyzers 330B, 330C, 330D, 331A, 332A, 333A, 334A, 339A, 8903A, 8903B, 8903E,???
- Dynamic measurement DC source
- Electrometers
- Fading simulators
- Femto ammeters
- Filters
- Frequency counters 522B, 5342A 5343A 5352B
- Frequency standards?
- Function Generators ? 3310A,? 8165A,
- GPIB controllers, extenders, cables etc.
- GPS frequency standards
- Harmonic mixers
- High resistance meters 4339B
- High resistance meter fixtures 16008B
- HEV EV Grid Emulators and Test Systems
- In-circuit test systems
- Impedance analyzers 4195A, 4291A, 4291B, 4395A, 4396A, 4396B, 4294A, E4990A, E4991A
- Impedance Analyzer Accessories
- Impedance / Gain Phase analyzer 4194A
- Impedance Meter 4193A,
- Isolators
- LCR meters? U1701A, U1701B, U1731A,? U1731B, U1731C, U1732A, U1732B, U1732C, U1733C, 4191A , 4192A, 4194A, 4195A, E4196A,? 4216A, 4260A, 4261A, 4262A? 4263A, 4263B, 4271B, 4274A, 4275A, 4276A , 4277A, 4284A, 4285A, 4286A, 4287A, 4291A, 4291B, 4294A, 4332A, 4342A, 4395A, 4396A, 4396B, E4980A and E4980AL
- LCR meter calibration devices? 16380A 42030A? 42090A, 42091A and 42100A
- LCR meter accessories
- 2-Terminal BNCs.
- 4-Terminal Pair (BNC connectors)
- Cable extension 16048A, 16048D, 16048E, 16048G, 16048H
- DC current bias accessories 42841A, 42842A, 42842B, 42842C, 42843A
- DC voltage bias accessories 16065A, 16065C,
- Kelvin clips 16089A, 16089B, 16089C,16089E
- Lead Components 16047A,16047B, 16047D, 16047E
- Material 16451B, 16452A
- Probes 42941A
- SMD 16034E, 16034G, 16034H
- 2-port 16096A
- 7 mm (APC7)
- 2-Terminal BNCs.
- LCZ meters? 4276A, 4277A,
- Lightwave clock / data receivers
- Lightwave converter
- Lightwave component analyzer
- Lightwave measurement system mainframes
- Lightwave polarization analyzers 8509B
- Logic analyzers
- Nemo wireless network solutions.
- Noise and interference test set
- Noise figure analyzers
- Noise sources 346A, 346B. 346C ,
- Matching pads (50 ohm to 75 ohm or similar)
- Materials test equipment
- Microwave repeaters
- Microwave downconverters 70427A
- Microwave / THz sources
- Milliammeter 428B
- Milliohm meter
- Mobile communications DC source
- Modular instruments
- AXIe
- Data acquisition (DAQ)
- USB
- PXIe
- Modulation analyzers
- Multimeters 427A, 970A
- Optical attenuators
- Optical heads
- Optical sources
- Optical spectrum analyzers
- Oscilloscopes 120A, 120AR, 120B, 122A, 130A, 130B, 130BR, 130C, 140A, 140B, 141A, 150A, 150AR, 160B, 180A, 180AR, 180CD, 181A, 181AR, 181T, 181TR, 182C, 182T, 183A, 183B, 184A, 184B, 185A, 185B, 1200A, 1200B, 1220A, 1221A, 1703A, 1707A, 1707B, 1710A, 1710B, 1715A, 1722A, 1725A, 1726A, 1740A, 1741A, 1742A, 1743A, 1744A, 1746A, 1980A, 1980B, 5403A, 6000A, 6000L, 16533A, 16534A, 54100A, 5410B, 54100C, 5100D, 54111D, 54120A, 54120B, 54200A, 54501A, 54502A, 54503A, 54504A, 54520A, 54520C, 54540A, 54540C, 54542A, 54542C, 54600B, 54601A, 54601B, 54602B, 54603B,? 54645A, 54654N, 54710A, 54720A, 54750A, 54825N, E1428,?
- Oven controlled crystal oscillators (OCXOs)
- Pattern generators
- PCM terminal test set
- Phase noise measurement
- Pico ammeters
- Printers 2225
- Plotters 7470A, 7475A?
- Probes
- Protocol analyzers and exercisers.
- Power booster test sets
- Power meters 431A, 431B, 431C, 432A, 435A, 435B, 437B, 438A
- Power splitters
- Power supplies
- Pulse generators
- Q-meters 4342A?
- Q-meter calibration inductors 16470A
- Reflection transmission test set
- Return loss module (optical)
- Relays / switches / switch matrices (optical)
- Relays / switches / switch matrices (RF)
- Resistor standards 42030A?and 42100A
- S-parameter test sets
- Scalar network analyzers
- SCSI bus preprocessor interface E2324A
- Selective level meters 3746A
- Semiconductors
- Semiconductor parameter analyzers 4145A, 4155B, 4156B,
- Signal analyzers
- Signal generators / sweep generators / signal sources / oscillators 200CD, 201B, 209A, 204D,? 608A,? 8165A
- Software
- Source measure units
- Spectrum analyzers 4195A,???
- Switch control units
- SWR meter 415E?
- Time interval? counters
- Time mark generator 226A
- Timing and data state modules
- Torque wrenches
- Transmitter testers
- Trigger modules
- Ultrasound transducers
- Universal bridge? 4260A, 4265A, 4265B?
- Vacuum tube voltmeter 410C
- Vector Impedance Meter 4193A, 4800A, 4815A
- Vector Network Analyzers (VNAs) 4195A,? 8510A, 8510B, 8510C, 8753A, 8753B, 8753C, 8753D, 8753E, 8753ES, 8752ET, 8719A, 8719B, 8719C, 8719D, 8720A, 8720B, 8720C, 8720D, 8720ES, 8722A, 8722B, 8722C, 8722D, 8722ES,
- Vector Network Analyzers (VNA) calibration kits 85032B, 85032E, 85033C, 85033D, 85033E, 85050B, 85050C, 85050D, 85052B, 85052C, 85052D, 85054A, 85054B, 85054D, 85056A
- Vector Network Analyzer (VNA) verification kits
- Vector Signal Analyzer 89650S, 89600S
- Vector voltmeters 8405A, 8508A,
- VXI mainframes 70000B, 70000C
- Waveform and function generators
- Waveguide to waveguide and waveguide to coaxial transitions.
- Wireless 58 OTA chambers
- Wireless channel emulators
- Wireless network emulators
- Wireless communication test sets
?
HP 54503A Memory address error
Hi All, Has anyone seen and fixed this error before I pull the case? Thanks Chris
By
Onewatt
·
#152052
·
|
Re: HP8510C "system bus address error", but then it works
Also check RF Switch address. It must be 31 if you havent switch. Wrong address cause "System bus error" message. [System][F2=HP-IB Addresses][F6=RF Switch]=31
By
§¥§Þ§Ú§ä§â§Ú§Û §À§â§î§Ö§Ó
·
#152051
·
|
Agilent/Keysight AWG 33250A /33220A CAL
hello everyone Bonjour a tous... This fine 80 MhZ AWG (circa 2016) needs a full calibration similar to the 33220A 20 MhZ. Do either of these generators, use a back up battery or NV ram that can lose
By
Froggie the Gremlin
·
#152050
·
|
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
And your reported 10 . 000447 MHz? results in error still.... Yes place your counter into ratio mode to properly address counter faults...
By
alan victor
·
#152049
·
|
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
I suggest you place your counter into the RATIO MODE.? What I am addressing is rule out a faulty divide by 2 counter.? ?Since this a significantly robust phase lock synthesizer system, makes sense
By
alan victor
·
#152048
·
|
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
That is pretty bad. An ovenized crystal oscillator should be able to do at least 10^-7. ¨C Mike Mike B. Feher, N4FS 89 Arnold Blvd. Howell NJ 07731 908-902-3831 Sent: Sunday, May 25, 2025 1:04 PM To:
By
Mike Feher
·
#152047
·
|
Re: Checking an 8402B Calibrator
It might be simpler to put a 100M ohm resistor in series with the high input on the 34401A in the High Z mode. Your 17.8V reading will now be 8.9V well inside the 10V limit for the 10V, 100M ohm
By
Don Bitters
·
#152046
·
|
Re: HP8510C "system bus address error", but then it works
?Marko, The system may have been set up to use two sources and it¡¯s looking for the second source. Try: Instrument State <Recall><More><Factory Preset> Regards, Steve
By
Steve - Home
·
#152045
·
|
HP8561E Jittery Display
Hi, I've some HP specans just obtained as BER items from a local space communication site. The first one I'm looking at is an HP8561E, it powers on OK and the display is bright but the text is jittery
By
Paul Randall G3NJV
·
#152044
·
|
HP8510C "system bus address error", but then it works
Got my hands on a 50GHz 8510C with test set and source, everything mounted in a mobile rack with a pullout table. When I switch it on, it says "caution: system bus address error", but them it proceeds
By
Marko Cebokli
·
#152043
·
|
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
*Actually measuring again with a counter I come up with 10.000447* *Not to say my equipment is perfectly accurate.*
By
Johnny10
·
#152042
·
|
Re: HP 8902A Measuring Receiver
Hey there Raymond, I've done a little futzing with the power supply and the input amp on my 8902, so I'm happy to try and help where I can. I think you might get more traction on the group if you
By
Tony Goodhew
·
#152041
·
|
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
Are you saying 4.999756 is in error?
By
Johnny10
·
#152040
·
|
Re: ftp.agilent.com tar-file on archive.org
I have been downloading files with "wget" for many years, but never used the "-c" option (I guess as long as one just downloads "regular" size files, it doesn't matter that much). It worked like a
By
Wolfgang Schraml
·
#152039
·
|
Re: Agilent E8357A Network Analyzer Phase Lock Lost Error
Why is the 5 MHz in error, while the 10 MHz is correct?
By
alan victor
·
#152038
·
|
Re: Checking an 8402B Calibrator
Tony, If you have or can borrow one, the Fluke 8840A does High-Z up to 20Vdc. --John Gord
By
John Gord
·
#152037
·
|
Re: Checking an 8402B Calibrator
Thanks, Jim & Charles - Appreciate it. Where I find myself is that the 8402B manual says to look at a voltage that is around 17.8 Vdc above the -30V rail (Section 5-7 in the manual) - This pushes my
By
Tony Goodhew
·
#152036
·
|
Re: Checking an 8402B Calibrator
Hi, the 34401A can do >10 G¦¸ input R but only on the 10V and below ranges: Blue Button (Menu) On/Off (Down arrow) (Right arrow) Enter See pages 27 and 53 of the manual (well the year 2000 version
By
Charles Edmonds
·
#152035
·
|
Re: Checking an 8402B Calibrator
Try using the 34401A in hi-Z (high impedance) mode, Tony. I had to do that once about 10 or 15 years ago when measuring voltage in a low current circuit. Got weird results until I switched it to hi-Z
By
Jim Ford
·
#152034
·
|
Checking an 8402B Calibrator
Hey All, I'm looking at an 8402B calibrator, and the manual says that the measurements need to be done with a floating DMM that has a greater than 100Mohm input impedance. The manual says to use an HP
By
Tony Goodhew
·
#152033
·
|