Re: E5071C powering down
"Sven Schnelle" <svens@...> writes:
By Sven Schnelle · #151096 ·
HP 6268B 0-40V 0-30A PS schematic
Anyone know where to find better than this scan from the Keysight site? It's not quite legible... https://www.ecosensory.com/tek/HP-6268B-op-svc-9018-05504-schem.png I found mil manual that is
By John Griessen · #151095 ·
HP 6268B 0-40V 0-30A PS schematic
Anyone know where to find better than this scan from the Keysight site? It's not quite legible... https://www.ecosensory.com/tek/HP-6268B-op-svc-9018-05504-schem.png
By John Griessen · #151094 ·
Re: HP6209B component location layout
You will also likely need to replace R12,13 parallel combination ~2K ohm 3W. R13 is a 1/2W trim resistor, this is between a 6.2V source and terminal A7 on the back. the equation is
By Hugh Gilbert · #151093 ·
Re: HP6209B component location layout
Thank you for the information, that saved me some ghost hunting for drifted resistors in the meter circuit. Both the voltage and current adjustment pots are Bourns series 3540 which are rated at
By Alex · #151092 ·
Re: LCR Meter E4980A Power On Test Fail
Sounds like that might be a firmware problem if it works when forced into an alternate mode. If there is a firmware download it might be worth checking. Regards Nigel
By nigel adams · #151091 ·
Re: Repair HP3455A Multimeter
Hi Orin, thank you for the hint,? i ordered 2N4149A from Dugikey, hope this will solve my cal #10 error. Ralf
By Ralf Rosche · #151090 ·
Re: HP6209B component location layout
Besides the effects on the metering, you may want to check the power rating of the volts pot. The original 100 kR value would dissipate about a watt at full output, now about 2 W with 50 kR. If these
By ed breya · #151089 ·
Re: Looking for the manual for the HP H04-3460B or 3460B opt H04
It would be better than nothing. I actually got rid of my fiche reader/printer about 8 years ago after not having used it in over 10 years. Never say never, LOL. Richard
By Richard Parrish · #151088 ·
Re: Repair HP3455A Multimeter
I'd try 2N4119A or PN4119A. I might have some... Mouser are mighty proud of them. [email protected]> wrote:
By Orin Eman · #151087 ·
Re: Repair HP3455A Multimeter
I don't find a 5070-7074 in the 3455A manual I downloaded, Manual P/N 03455-90003. Looking for Q19, I find A10Q3, 4, 15, 16, 18, 19, 21? which is a 5081-7074 and A15Q19, 20, which is a
By John Kolb · #151086 ·
HP 432 Replacement
Hi All Has anyone used the TEGEM ?1830A RF Power Meter? a? HP 432 Replacement or the new Keysight Replacement for the HP 432 as i have just bought a HP 478A? Option H72 and would like to get an
By Paul Bicknell · #151085 ·
Re: Looking for the manual for the HP H04-3460B or 3460B opt H04
ArtekManuals have at least one version of the 3460A manual, but not the H04 option. I also have some info on the chopper oscillator modification, used in the later 3459/60 voltmeters, which is
By factory · #151084 ·
Re: Looking for the manual for the HP H04-3460B or 3460B opt H04
The H04 manual for the 3460 *A* is just the normal manual with some change sheets for the H04 modifications, if there is a 3460 *B* H04 version you should be able to use the 3460 *B* manual at the web
By factory · #151083 ·
Re: Looking for the manual for the HP H04-3460B or 3460B opt H04
I have some info for the 3460 *A* -H04 but not the B Looks like specs / service information. It is part of a set of fiches that cover 3460A specials and is a little randomly organized and poorly
By Peter Brown · #151082 ·
Re: Repair HP3455A Multimeter
Ralf, Unless I am very mucht mistaken (...) the 3325 uses the chassis as GND, and **all** screws fastening the PCB should better be nice and tight. The PCB has no overall GND itself. This can drive
By Wilko Bulte · #151081 ·
Looking for the manual for the HP H04-3460B or 3460B opt H04
I'm looking for the Service manual for the H04 version of the HP 3460B. H04 adds an additional digit to the display and has improved accuracy. Richard
By Richard Parrish · #151080 ·
Re: HP6209B component location layout
The current through the 100k pot is normally 3.2mA.to change that to 50k would require doubling the current to 6.4mA. This additional current will show up as an offset on the current meter. The 6207
By Hugh Gilbert · #151079 ·
HP6209B component location layout
I recently acquired a HP6209B, and already have the same manual as the one here in the files section, but it does not include the component layout, as it seems that page was left out during the scan.
By Alex · #151078 ·
LCR Meter E4980A Power On Test Fail
Hi, I am currently trying to repair an E4980A LCR Meter that is experiencing a power-on failure. When the unit is powered on, the self-test displays the following error message: *"E1000: Power On
By KC Chong · #151077 ·