CHECK OUT THE WIKI The purpose of the wiki is mainly to allow you to find information on instruments, either from either
- The model number
- The function(s) listed below. Some instruments have multiple functions - for example, the 4195A is a VNA, spectrum analyzer and an impedance analyzer. Therefore the 4195A is listed in multiple categories
Please also check out HPWiki available here:
- Accessory kits - various types
- AC power analyzers - PA2201A and PA2203A
- AC power supplies 6811C, 6812C, 6813C
- Airlines
- Arbitrary waveform generators M8194A
- Amplifiers?493A, 495A?
- Attenuators (optical) 8156A, 8157A, 8158B, 81566A, 81576A,?
- Attenuators (RF) 8494A
- Attenuator set (500 Ω) 350C
- Attenuator set (600 Ω) 350D
- Attenuator switch driver
- Audio analyzers? 8903A, 8903B, 8903E,? ?
- Base station test sets
- Bit error rate testers (BERTs)
- Cables
- Capacitance meters U1701A, U1701B, 4272A, 4278A, 4279A
- Capacitor Bridge 4270A,
- Capacitor standards 16380A, 16380C,?
- Carrier noise test setsi
- Cesium frequency standards
- Clamp ammeters
- Close field probes
- Crystal Impedance E4915A, E4916A
- Data Acquisition Systems (DAQs)
- DC power analyzers
- DC power supplies 6030A , 6031A , 6032A, 6033A, 6035A, 6131C, 6621A, 6622A, 6623A, 6624A, 6627A, 6255A, 6645A, 6671A, 6672A, 6673A, 6674A, 6675A, 62003A, 62003C, 62003E, 62004A, 62004B, 62004E, 62005A, 62005B, 62005E, 62006A, 62006B, 62006E, 62010A, 62010C, 62010E, 62012A, 62012C, 62012E, 62015A, 62015C, 62015E, 62018A, 62018C, 62018E, 62024A, 62024C, 62024E, 62028A, 62028C, 62028E, 62048A, 62048C, 62048E
- Delay lines
- Detectors
- Device current waveform analyzers
- Digital communications analyzers
- Directional couplers
- Distortion analyzers 330B, 330C, 330D, 331A, 332A, 333A, 334A, 339A, 8903A, 8903B, 8903E,???
- Dynamic measurement DC source
- Electrometers
- Fading simulators
- Femto ammeters
- Filters
- Frequency counters 522B, 5342A 5343A 5352B
- Frequency standards?
- Function Generators ? 3310A,? 8165A,
- GPIB controllers, extenders, cables etc.
- GPS frequency standards
- Harmonic mixers
- High resistance meters 4339B
- High resistance meter fixtures 16008B
- HEV EV Grid Emulators and Test Systems
- In-circuit test systems
- Impedance analyzers 4195A, 4291A, 4291B, 4395A, 4396A, 4396B, 4294A, E4990A, E4991A
- Impedance Analyzer Accessories
- Impedance / Gain Phase analyzer 4194A
- Impedance Meter 4193A,
- Isolators
- LCR meters? U1701A, U1701B, U1731A,? U1731B, U1731C, U1732A, U1732B, U1732C, U1733C, 4191A , 4192A, 4194A, 4195A, E4196A,? 4216A, 4260A, 4261A, 4262A? 4263A, 4263B, 4271B, 4274A, 4275A, 4276A , 4277A, 4284A, 4285A, 4286A, 4287A, 4291A, 4291B, 4294A, 4332A, 4342A, 4395A, 4396A, 4396B, E4980A and E4980AL
- LCR meter calibration devices? 16380A 42030A? 42090A, 42091A and 42100A
- LCR meter accessories
- 2-Terminal BNCs.
- 4-Terminal Pair (BNC connectors)
- Cable extension 16048A, 16048D, 16048E, 16048G, 16048H
- DC current bias accessories 42841A, 42842A, 42842B, 42842C, 42843A
- DC voltage bias accessories 16065A, 16065C,
- Kelvin clips 16089A, 16089B, 16089C,16089E
- Lead Components 16047A,16047B, 16047D, 16047E
- Material 16451B, 16452A
- Probes 42941A
- SMD 16034E, 16034G, 16034H
- 2-port 16096A
- 7 mm (APC7)
- 2-Terminal BNCs.
- LCZ meters? 4276A, 4277A,
- Lightwave clock / data receivers
- Lightwave converter
- Lightwave component analyzer
- Lightwave measurement system mainframes
- Lightwave polarization analyzers 8509B
- Logic analyzers
- Nemo wireless network solutions.
- Noise and interference test set
- Noise figure analyzers
- Noise sources 346A, 346B. 346C ,
- Matching pads (50 ohm to 75 ohm or similar)
- Materials test equipment
- Microwave repeaters
- Microwave downconverters 70427A
- Microwave / THz sources
- Milliammeter 428B
- Milliohm meter
- Mobile communications DC source
- Modular instruments
- AXIe
- Data acquisition (DAQ)
- USB
- PXIe
- Modulation analyzers
- Multimeters 427A, 970A
- Optical attenuators
- Optical heads
- Optical sources
- Optical spectrum analyzers
- Oscilloscopes 120A, 120AR, 120B, 122A, 130A, 130B, 130BR, 130C, 140A, 140B, 141A, 150A, 150AR, 160B, 180A, 180AR, 180CD, 181A, 181AR, 181T, 181TR, 182C, 182T, 183A, 183B, 184A, 184B, 185A, 185B, 1200A, 1200B, 1220A, 1221A, 1703A, 1707A, 1707B, 1710A, 1710B, 1715A, 1722A, 1725A, 1726A, 1740A, 1741A, 1742A, 1743A, 1744A, 1746A, 1980A, 1980B, 5403A, 6000A, 6000L, 16533A, 16534A, 54100A, 5410B, 54100C, 5100D, 54111D, 54120A, 54120B, 54200A, 54501A, 54502A, 54503A, 54504A, 54520A, 54520C, 54540A, 54540C, 54542A, 54542C, 54600B, 54601A, 54601B, 54602B, 54603B,? 54645A, 54654N, 54710A, 54720A, 54750A, 54825N, E1428,?
- Oven controlled crystal oscillators (OCXOs)
- Pattern generators
- PCM terminal test set
- Phase noise measurement
- Pico ammeters
- Printers 2225
- Plotters 7470A, 7475A?
- Probes
- Protocol analyzers and exercisers.
- Power booster test sets
- Power meters 431A, 431B, 431C, 432A, 435A, 435B, 437B, 438A
- Power splitters
- Power supplies
- Pulse generators
- Q-meters 4342A?
- Q-meter calibration inductors 16470A
- Reflection transmission test set
- Return loss module (optical)
- Relays / switches / switch matrices (optical)
- Relays / switches / switch matrices (RF)
- Resistor standards 42030A?and 42100A
- S-parameter test sets
- Scalar network analyzers
- SCSI bus preprocessor interface E2324A
- Selective level meters 3746A
- Semiconductors
- Semiconductor parameter analyzers 4145A, 4155B, 4156B,
- Signal analyzers
- Signal generators / sweep generators / signal sources / oscillators 200CD, 201B, 209A, 204D,? 608A,? 8165A
- Software
- Source measure units
- Spectrum analyzers 4195A,???
- Switch control units
- SWR meter 415E?
- Time interval? counters
- Time mark generator 226A
- Timing and data state modules
- Torque wrenches
- Transmitter testers
- Trigger modules
- Ultrasound transducers
- Universal bridge? 4260A, 4265A, 4265B?
- Vacuum tube voltmeter 410C
- Vector Impedance Meter 4193A, 4800A, 4815A
- Vector Network Analyzers (VNAs) 4195A,? 8510A, 8510B, 8510C, 8753A, 8753B, 8753C, 8753D, 8753E, 8753ES, 8752ET, 8719A, 8719B, 8719C, 8719D, 8720A, 8720B, 8720C, 8720D, 8720ES, 8722A, 8722B, 8722C, 8722D, 8722ES,
- Vector Network Analyzers (VNA) calibration kits 85032B, 85032E, 85033C, 85033D, 85033E, 85050B, 85050C, 85050D, 85052B, 85052C, 85052D, 85054A, 85054B, 85054D, 85056A
- Vector Network Analyzer (VNA) verification kits
- Vector Signal Analyzer 89650S, 89600S
- Vector voltmeters 8405A, 8508A,
- VXI mainframes 70000B, 70000C
- Waveform and function generators
- Waveguide to waveguide and waveguide to coaxial transitions.
- Wireless 58 OTA chambers
- Wireless channel emulators
- Wireless network emulators
- Wireless communication test sets
?
Re: Service manual scan post processing
Hi Tom,
Not a digital camera a Canon MS-800 microfiche scanner that was being e-cycled by a local engineering firm
Canon MS800II Microfilm Scanner | The Scanner Shop (
By
Peter Brown
·
#150233
·
|
Re: Service manual scan post processing
[email protected]> wrote:
If I understand correctly, you are using a digital camera to take a
picture, then processing that.
Could you please upload the jpg file produced by the camera, without
By
Tom Gardner
·
#150232
·
|
Re: Service manual scan post processing
Hi David,
I am wondering if the claimed 600 dpi is actually the result of interpolation / dithering
I will upload a page at 600 DPI in bmp format for completeness in case the no compression .tiff
By
Peter Brown
·
#150231
·
|
Re: Service manual scan post processing
<tggzzz@...> wrote:
Thanks, I'll have to try and remember that script in the future.
My current process converts things to PNM, iterates over 0.0 to 0.9
(by 0.1) into pamthreshold,
By
David Holland
·
#150230
·
|
Re: Service manual scan post processing
Nope, those appear to be dithered too.. (and show indications of JPEG
compression).
I uploaded a high zoom'd screenshot of what I'm seeing..
Sorry...
<peter@...> wrote:
By
David Holland
·
#150229
·
|
Re: VNA Fixtures
That may well be necessary if I encounter one of those 'problem child' devices that Ed described!
By
Jinxie
·
#150228
·
|
Re: Service manual scan post processing
Hi David
I have added some of the raw .tiff output from the MS-800 in a tiff subdirectory.
Perhaps these are more suitable for contrast enhance etc?
Peter
By
Peter Brown
·
#150227
·
|
Re: Service manual scan post processing
[email protected]> wrote:
That is highly beneficial, and is the way the scancvt script I posted
earlier reduces pages to ~80kBytes for a page of text.
The scancvt script determines the average
By
Tom Gardner
·
#150226
·
|
Re: VNA Fixtures
If you have a lot you can carefully disassemble one of each to get further knowledge.
Peter
By
Peter Gottlieb
·
#150225
·
|
Re: VNA Fixtures
Ed, many thanks for that comprehensive answer. All bar one of these devices have that capacitor you describe tucked away in the base. Can't be more than a few tens of pF at most I'd have thought. The
By
Jinxie
·
#150224
·
|
Re: Service manual scan post processing
I think you may be missing a contrast maximization step. If you zoom
in, the pages appear to be greyscale, dithered down to black and
white. I don't think you're going to get much better compression
By
David Holland
·
#150223
·
|
Re: Disassembling HP 64000 code
Thank you Don and Steve. The assembly code looks exactly like how FORTH is
described. The main code sets up a processing engine which gets fed lists
of pointers that reference either real assemble
By
Tom Haynes
·
#150222
·
|
Re: Dissassembling HP 65000 code
Sven,
Can you reply to the new thread with the Bitsaver and Mame information?
Both are valuable to the discussion and I hope to include it in the correct
thread.
Thanks
[email protected]>
By
Tom Haynes
·
#150221
·
|
Re: Service manual scan post processing
Hi Martin,
Thanks for the steer on ImageOptim, I will take a look.
I am in touch with a service engineer who used to run a company that printed microfiche and also supplied the reader / scanners.
He
By
Peter Brown
·
#150220
·
|
Re: Service manual scan post processing
Hi Michael,
Many thanks for responding.? Please do take a look at the files, I have posted some pdfs here
*Files - A temporary directory for photographs and help relating to emails and posting -
By
Peter Brown
·
#150219
·
|
Re: HP 8670B amplifier - does it ring a bell with anyone?
If System Component serial numbers follow classic Hewlett Packard rules, that 6870B would have been built in 1990.? That said, there seem to be 1991 and 1992 stampings on the rear upper left of the
By
Bert
·
#150218
·
|
Re: Disassembling HP 64000 code
I was also thinking from your description of the memory contents you found, that it sounds like a stored FORTH program. I implemented FORTH-77 on a 6502 back in the early 80's, and I still have some
By
Steve Hendrix
·
#150217
·
|
Re: HP8673C Restoration project
Allan thats a nice generator. I did troubleshoot one for a friend 5 years
ago. It turned out to be the yig coil and no parts were available. He did
sell the generator as is and did fine. But that was
By
paulswed
·
#150216
·
|
Re: Disassembling HP 64000 code
You might want to check this site for information about FORTH. I believe the HP64000 did support FORTH in some of the early versions.
What is the Forth programming language?
forth.com
Don Bitters
By
Don Bitters
·
#150215
·
|
Re: HP 8670B amplifier - does it ring a bell with anyone?
There is also an 8760B K73 which is a 24 input 2 output switch matrix.
The unit is 19" 4U in size.
I am not sure for what systems were these used.
By
Razvan Popescu
·
#150214
·
|