¿ªÔÆÌåÓý

Date
Re: HP141T with multiple sections
You can turn on the 141 display by itself, but will only have a dot display which you can move with the horiz/vert controls. I would not leave it on for any length of time with no plug-ins installed.
By Don Bitters · #148853 ·
Re: Fan replacement 8341B
Yes, it is quiet, hard to tell it is on. I should have noted, I used a fan that has slightly less air flow than the original. This instrument will be used on the bench, at room temperature, not
By Gary Hitchner · #148852 ·
Re: Fan replacement 8341B
Good job, Is it any quieter than the original?
By saevar · #148851 ·
Re: HP141T with multiple sections
This depends on the age of the mainframe, late ones have a interlock connection & relay in the power supply, to prevent operation with no plug-ins fitted. But for testing something in unknown
By factory · #148850 ·
Re: Looking for Noise Source Driver Assembly E4401-60123 for E4407B Spectrum Analyzer
I'm afraid I've never looked inside any of the NFA units. They do look like an ESA on the outside but I think the internal signal path will have quite a few changes. The noise figure of the NFA will
By jmr · #148849 ·
Re: 8350A + 83525A mechanical issues
I guess I should say IF there's a battery problem. The message flow is a little confusing to me - not sure about any battery issues.?? Ed
By ed breya · #148848 ·
Re: 8350A + 83525A mechanical issues
Since you have a battery leakage problem too I'd recommend cleanup, repair, and checkout of proper operation before wrapping up the switch and connector issues. If there's fatal damage from the
By ed breya · #148847 ·
Fan replacement 8341B
Posting for information if anybody might be able to use it. My fan replacement in the 8341B was successful. The Orion fan number OA172SAP-11-3TB is just about an exact replacement. Two differences
By Gary Hitchner · #148846 ·
Re: Shielded probe cables?
If you're going to do an adapter, according to the user manual the 5032 has an input resistance of 220k and an input capacitance of 12pF.? You could adjust the 10k termination up a little (10.476k)
By Mark Litwack · #148845 ·
Re: HP 8341B 1992 firmware version
I don¡¯t suppose you¡¯d be willing to write out a copy of the 8.xx image from the device to a floppy and upload it would you? Or if you have files, upload those? I don¡¯t see anything other than
By Ed Marciniak · #148844 ·
Re: 8350A + 83525A mechanical issues
Hi Bumbo! Where are you located?J.Kruth In a message dated 12/3/2024 4:25:12 AM Eastern Standard Time, bumbo.chinelo@... writes:?Thanks for the feedback?Last night I tried the rocker
By Jeff Kruth · #148843 ·
Re: HP141T with multiple sections
Hello, In France you can place a ad on : https://radioamateur.org/ There is one actually for sale - pick-up only - for 400 Euros ( HP141T Mainframe + 8552B , 8553B and 8555A ) Also on :
By f1ghb · #148842 ·
Re: HP141T with multiple sections
If you need a hand to deliver it to the junk yard ........ /s/s
By bumbo.chinelo@... · #148841 ·
Re: HP141T with multiple sections
Hello Albert, before the junkyard I'd put it on Ebay, 1 Euro starting price, pickup only. For sure there will be some bids! a+ Martin
By Martin · #148840 ·
HP141T with multiple sections
Hello, I'm the owner of a HP141T with multiple sections since 2010. From that time, in order to be quick on different projects and testing, I progressively purchased three HP8552B? sections (IF
By tantratron · #148839 ·
Re: 8350A + 83525A mechanical issues
Thanks for the feedback Last night I tried the rocker fix. It almost worked, but when I went to put the final turns on the screws the plastic cracked thus I went to the next task, i.e. kicking myself
By bumbo.chinelo@... · #148838 ·
Re: Looking for Noise Source Driver Assembly E4401-60123 for E4407B Spectrum Analyzer
Thanks for the PDF, very interesting reading. Do you know what is the difference between the NFA and the ESA? I thought the NFA is just an ESA with the Option 219 card installed and the 2 ports routed
By Razvan Popescu · #148837 ·
Re: Looking for Noise Source Driver Assembly E4401-60123 for E4407B Spectrum Analyzer
Hi, there is a similar graph of uncertainties, but with VERY different axis scales, in 5989-8056EN dated May 12 2008. It also shows that the NF option for the ESA analyzers to be less than wonderful.
By Charles Edmonds · #148836 ·
Re: HP 8341B 1992 firmware version
Reasonably sure you will not have a problem - please do report the result. Cheers! Bruce
By Bruce · #148835 ·
Re: Looking for Noise Source Driver Assembly E4401-60123 for E4407B Spectrum Analyzer
The E444xA does have the 110 option for a 50 GHz preamp but the noise figure (and presumably also the input VSWR) of the preamp limits the performance across the upper parts of the frequency range.
By jmr · #148834 ·