Re: HP 339A Distortion meter problem
I had an issue with my 339, where it would work in some switch positions but not others, and even depending on which direction the switches were moved (probably why I scored it so cheap!). I traced it
By romeo987 · #148590 ·
Re: Troubleshooting a 8591e with opt 130
Yves, Agreed, I won't touch any adjustment yet. Wouldn't surprise me though if a previous owner already did. Anyway it seems we are locking in on the issue thanks to all your help which is great. I
By Marco · #148589 ·
Re: Troubleshooting a 8591e with opt 130
Marco, Obviously, it seems that it is the LOG section that is lacking, since the LIN mode works with the correct RF level. It may be one of the gain controls in module A. I will look at the behavior
By Yves Tardif · #148588 ·
Re: Troubleshooting a 8591e with opt 130
Hey Gordon, Thanks for the additional info! For now I'll keep trying to fix it with this firmware installed, I like the HP logo more anyway. ;) Marco
By Marco · #148587 ·
Re: HP 8662A + 8663A Power Supply
My 8663 was a flea market "bargain" (well - not really) but it was repairable by replacing the inverter transistors with BUX48 as well as a bunch of electrolytic capacitors and some other components
By Ulf Kylenfall · #148586 ·
Re: Troubleshooting a 8591e with opt 130
HI Marco I have two HP8591E spectrum analysers one has several options installed including opt 130 the other had no option installed. I found opt 130 and opt 004 on eBay I just copied the wiring from
By Gordon Slessor · #148585 ·
Re: HP 8662A + 8663A Power Supply
Hi Dan I would be interested in purchasing a set of spare boards. I am located in Australia zip 3095z. My unit 8663a has been working perfectly but the original power supply has taken out the
By tekfreak09 · #148584 ·
Re: Troubleshooting a 8591e with opt 130
So considering all the RBWs with the Log path are off, should I search for a fault in the A14 Log amplifier Assy? Or do you still think it's somewhere in the filters? I could try to do the 7. Log and
By Marco · #148583 ·
Re: HP 339A Distortion meter problem
MG has some interesting alternatives, some even more expensive than the Deoxit offerings (!!). For instance, the 801B. I still rely mostly on the Deoxit, but in some cases (high impedance
By Radu Bogdan Dicher · #148582 ·
Re: Connector identification
Hi Scott, I think you solved the mystery. OSMT connector has the right shape and dimensions. Ozan
By Ozan · #148581 ·
Re: HP 339A Distortion meter problem
I was going to recommend De-Oxit as well - I haven't found any other cleaner that works as well, which seems to justify its higher-than-I-think-it-should-be price.? I usually use De-Oxit Gold, except
By Mark Moulding · #148580 ·
Re: HP8920A, No receiving and when ROM Self test runed, it show 11 fails in the Receiver synthesizer and when the receiver test runed also shows 11 fails.
/g/HP-Agilent-Keysight-equipment/files/All%20HP,%20Agilent%20and%20Keysight%20instruments%20in%20folders%20by%20part%20numbers/8000%20to%208999/8920-8921
By Mahlon Haunschild - N4ZK · #148579 ·
Re: HP8920A, No receiving and when ROM Self test runed, it show 11 fails in the Receiver synthesizer and when the receiver test runed also shows 11 fails.
The CLIPS are in this group¡¯s files section. ?They are not on the Keysight website. best regards Mahlon - N4ZK
By Mahlon Haunschild - N4ZK · #148578 ·
Re: Troubleshooting a 8591e with opt 130
Since the picture you send about BW5 etc went down to 300Hz, I assumed those were the standard RBW settings. But it sounds more plausible anyway that all the standard RBW have the issue and the ones
By Marco · #148577 ·
Re: Troubleshooting a 8591e with opt 130
Without option 130, the last resBW is 1kHz. With option 130: 300,200,100, and 30 Hz There a problem with standard resBW. De : [email protected]
By Yves Tardif · #148576 ·
Re: Troubleshooting a 8591e with opt 130
Marco, Both PCBs have the same part number. But in my case, the A13 label on the PCB was for the right location i.e. A13...A11 PCB with A11 location. So, I don't see any difference between the PCBs,
By Yves Tardif · #148575 ·
Re: N8973A Firmware Update issue
Hi Vladan, thanks for the reply and the offered help. Today I was able to create a new set of disks on an old XP based system following Antoine¡®s hint. I will check this on Saturday as I am not at
By DL8EBL · #148574 ·
Re: HP8920A, No receiving and when ROM Self test runed, it show 11 fails in the Receiver synthesizer and when the receiver test runed also shows 11 fails.
You will need to troubleshoot the synthesizer. Start downloading the CLIP from Keysight site. Ing. Patricio A. Greco Taller Aeron¨¢utico de Reparaci¨®n 1B-349 Organizaci¨®n de Mantenimiento
By Patricio A. Greco · #148573 ·
Re: HP8920A, No receiving and when ROM Self test runed, it show 11 fails in the Receiver synthesizer and when the receiver test runed also shows 11 fails.
Hi Patricio yes, is no locked, I replaced the receiver synth board and I got the same result. Should be one of the signal coming to the synth, The reference test is ok. Do you have any advice? If you
By Sebastian Fernandez · #148572 ·
Re: Troubleshooting a 8591e with opt 130
I did a reseating of A13 and A11. Again found something interesting. The manual states A13 and A11 are identical boards but A13 is the one nearest to the PSU. In my analyzer A13 and A11 is written on
By Marco · #148571 ·