CHECK OUT THE WIKI The purpose of the wiki is mainly to allow you to find information on instruments, either from either
- The model number
- The function(s) listed below. Some instruments have multiple functions - for example, the 4195A is a VNA, spectrum analyzer and an impedance analyzer. Therefore the 4195A is listed in multiple categories
Please also check out HPWiki available here:
- Accessory kits - various types
- AC power analyzers - PA2201A and PA2203A
- AC power supplies 6811C, 6812C, 6813C
- Airlines
- Arbitrary waveform generators M8194A
- Amplifiers?493A, 495A?
- Attenuators (optical) 8156A, 8157A, 8158B, 81566A, 81576A,?
- Attenuators (RF) 8494A
- Attenuator set (500 ¦¸) 350C
- Attenuator set (600 ¦¸) 350D
- Attenuator switch driver
- Audio analyzers? 8903A, 8903B, 8903E,? ?
- Base station test sets
- Bit error rate testers (BERTs)
- Cables
- Capacitance meters U1701A, U1701B, 4272A, 4278A, 4279A
- Capacitor Bridge 4270A,
- Capacitor standards 16380A, 16380C,?
- Carrier noise test setsi
- Cesium frequency standards
- Clamp ammeters
- Close field probes
- Crystal Impedance E4915A, E4916A
- Data Acquisition Systems (DAQs)
- DC power analyzers
- DC power supplies 6030A , 6031A , 6032A, 6033A, 6035A, 6131C, 6621A, 6622A, 6623A, 6624A, 6627A, 6255A, 6645A, 6671A, 6672A, 6673A, 6674A, 6675A, 62003A, 62003C, 62003E, 62004A, 62004B, 62004E, 62005A, 62005B, 62005E, 62006A, 62006B, 62006E, 62010A, 62010C, 62010E, 62012A, 62012C, 62012E, 62015A, 62015C, 62015E, 62018A, 62018C, 62018E, 62024A, 62024C, 62024E, 62028A, 62028C, 62028E, 62048A, 62048C, 62048E
- Delay lines
- Detectors
- Device current waveform analyzers
- Digital communications analyzers
- Directional couplers
- Distortion analyzers 330B, 330C, 330D, 331A, 332A, 333A, 334A, 339A, 8903A, 8903B, 8903E,???
- Dynamic measurement DC source
- Electrometers
- Fading simulators
- Femto ammeters
- Filters
- Frequency counters 522B, 5342A 5343A 5352B
- Frequency standards?
- Function Generators ? 3310A,? 8165A,
- GPIB controllers, extenders, cables etc.
- GPS frequency standards
- Harmonic mixers
- High resistance meters 4339B
- High resistance meter fixtures 16008B
- HEV EV Grid Emulators and Test Systems
- In-circuit test systems
- Impedance analyzers 4195A, 4291A, 4291B, 4395A, 4396A, 4396B, 4294A, E4990A, E4991A
- Impedance Analyzer Accessories
- Impedance / Gain Phase analyzer 4194A
- Impedance Meter 4193A,
- Isolators
- LCR meters? U1701A, U1701B, U1731A,? U1731B, U1731C, U1732A, U1732B, U1732C, U1733C, 4191A , 4192A, 4194A, 4195A, E4196A,? 4216A, 4260A, 4261A, 4262A? 4263A, 4263B, 4271B, 4274A, 4275A, 4276A , 4277A, 4284A, 4285A, 4286A, 4287A, 4291A, 4291B, 4294A, 4332A, 4342A, 4395A, 4396A, 4396B, E4980A and E4980AL
- LCR meter calibration devices? 16380A 42030A? 42090A, 42091A and 42100A
- LCR meter accessories
- 2-Terminal BNCs.
- 4-Terminal Pair (BNC connectors)
- Cable extension 16048A, 16048D, 16048E, 16048G, 16048H
- DC current bias accessories 42841A, 42842A, 42842B, 42842C, 42843A
- DC voltage bias accessories 16065A, 16065C,
- Kelvin clips 16089A, 16089B, 16089C,16089E
- Lead Components 16047A,16047B, 16047D, 16047E
- Material 16451B, 16452A
- Probes 42941A
- SMD 16034E, 16034G, 16034H
- 2-port 16096A
- 7 mm (APC7)
- 2-Terminal BNCs.
- LCZ meters? 4276A, 4277A,
- Lightwave clock / data receivers
- Lightwave converter
- Lightwave component analyzer
- Lightwave measurement system mainframes
- Lightwave polarization analyzers 8509B
- Logic analyzers
- Nemo wireless network solutions.
- Noise and interference test set
- Noise figure analyzers
- Noise sources 346A, 346B. 346C ,
- Matching pads (50 ohm to 75 ohm or similar)
- Materials test equipment
- Microwave repeaters
- Microwave downconverters 70427A
- Microwave / THz sources
- Milliammeter 428B
- Milliohm meter
- Mobile communications DC source
- Modular instruments
- AXIe
- Data acquisition (DAQ)
- USB
- PXIe
- Modulation analyzers
- Multimeters 427A, 970A
- Optical attenuators
- Optical heads
- Optical sources
- Optical spectrum analyzers
- Oscilloscopes 120A, 120AR, 120B, 122A, 130A, 130B, 130BR, 130C, 140A, 140B, 141A, 150A, 150AR, 160B, 180A, 180AR, 180CD, 181A, 181AR, 181T, 181TR, 182C, 182T, 183A, 183B, 184A, 184B, 185A, 185B, 1200A, 1200B, 1220A, 1221A, 1703A, 1707A, 1707B, 1710A, 1710B, 1715A, 1722A, 1725A, 1726A, 1740A, 1741A, 1742A, 1743A, 1744A, 1746A, 1980A, 1980B, 5403A, 6000A, 6000L, 16533A, 16534A, 54100A, 5410B, 54100C, 5100D, 54111D, 54120A, 54120B, 54200A, 54501A, 54502A, 54503A, 54504A, 54520A, 54520C, 54540A, 54540C, 54542A, 54542C, 54600B, 54601A, 54601B, 54602B, 54603B,? 54645A, 54654N, 54710A, 54720A, 54750A, 54825N, E1428,?
- Oven controlled crystal oscillators (OCXOs)
- Pattern generators
- PCM terminal test set
- Phase noise measurement
- Pico ammeters
- Printers 2225
- Plotters 7470A, 7475A?
- Probes
- Protocol analyzers and exercisers.
- Power booster test sets
- Power meters 431A, 431B, 431C, 432A, 435A, 435B, 437B, 438A
- Power splitters
- Power supplies
- Pulse generators
- Q-meters 4342A?
- Q-meter calibration inductors 16470A
- Reflection transmission test set
- Return loss module (optical)
- Relays / switches / switch matrices (optical)
- Relays / switches / switch matrices (RF)
- Resistor standards 42030A?and 42100A
- S-parameter test sets
- Scalar network analyzers
- SCSI bus preprocessor interface E2324A
- Selective level meters 3746A
- Semiconductors
- Semiconductor parameter analyzers 4145A, 4155B, 4156B,
- Signal analyzers
- Signal generators / sweep generators / signal sources / oscillators 200CD, 201B, 209A, 204D,? 608A,? 8165A
- Software
- Source measure units
- Spectrum analyzers 4195A,???
- Switch control units
- SWR meter 415E?
- Time interval? counters
- Time mark generator 226A
- Timing and data state modules
- Torque wrenches
- Transmitter testers
- Trigger modules
- Ultrasound transducers
- Universal bridge? 4260A, 4265A, 4265B?
- Vacuum tube voltmeter 410C
- Vector Impedance Meter 4193A, 4800A, 4815A
- Vector Network Analyzers (VNAs) 4195A,? 8510A, 8510B, 8510C, 8753A, 8753B, 8753C, 8753D, 8753E, 8753ES, 8752ET, 8719A, 8719B, 8719C, 8719D, 8720A, 8720B, 8720C, 8720D, 8720ES, 8722A, 8722B, 8722C, 8722D, 8722ES,
- Vector Network Analyzers (VNA) calibration kits 85032B, 85032E, 85033C, 85033D, 85033E, 85050B, 85050C, 85050D, 85052B, 85052C, 85052D, 85054A, 85054B, 85054D, 85056A
- Vector Network Analyzer (VNA) verification kits
- Vector Signal Analyzer 89650S, 89600S
- Vector voltmeters 8405A, 8508A,
- VXI mainframes 70000B, 70000C
- Waveform and function generators
- Waveguide to waveguide and waveguide to coaxial transitions.
- Wireless 58 OTA chambers
- Wireless channel emulators
- Wireless network emulators
- Wireless communication test sets
?
Re: Anyone able to do GC or GC-MS on Hp ThinkJet Black ink ( for the Hp ThinkJet 2225 printer; Hp 9226a Or Hp 51604a ink cartridges ) ?
Don't think I need an outside lab, .... but would be interested in discuss with those actually having analytical equipment.
(? I still need a few minor parts for my pe Elan 6000,? so that is not
By
gren
·
#147709
·
|
Re: Anyone able to do GC or GC-MS on Hp ThinkJet Black ink ( for the Hp ThinkJet 2225 printer; Hp 9226a Or Hp 51604a ink cartridges ) ?
It could be that Food Black 2 is used but what they have made confidential is the salt that they use to neutralize the dye.? In this case Food black 2 has 4 sulfonic acid groups that need to be
By
Frank Mashockie
·
#147708
·
|
Re: Boonton 92BD Troubleshooting
Brought a 260 q meter back to life earlier this year. Had some idea of
what happened with them and hp, but not that detailed. Thanks for that!
:)
[email protected]> wrote:
By
Bill Berzinskas
·
#147707
·
|
Re: Boonton 92BD Troubleshooting
FWIW, I was one of those who promoted the Boonton list. There were two Boonton companies: Boonton Radio, who made the Q-Meter, RX-Meter and a series of signal generators, and Boonton Electronics
By
Richard Knoppow
·
#147706
·
|
Re: Anyone able to do GC or GC-MS on Hp ThinkJet Black ink ( for the Hp ThinkJet 2225 printer; Hp 9226a Or Hp 51604a ink cartridges ) ?
Well, RIT dye contains salt, and what dyes doesn't
FDA think are food dyes? They allow Red #40, and
yellow #5...
It is said that if you hear the sound of hoof beats
on the commons, it is more
By
Chuck Harris
·
#147705
·
|
Re: Boonton 92BD Troubleshooting
I personally don't mind your post. I just mentioned the other group in case you were unaware of it.
DaveD
KC0WJN
Thanks for all the fish.
==============================
All spelling mistakes are the
By
Dave Daniel
·
#147704
·
|
Re: Boonton 92BD Troubleshooting
It looked very underloved but if that's not the case I can move this post.
[email protected]> wrote:
By
Bill Berzinskas
·
#147703
·
|
Re: Boonton 92BD Troubleshooting
There is a Boonton group.
DaveD
KC0WJN
Thanks for all the fish.
==============================
All spelling mistakes are the responsibilty of the reader (Rick Renz, STK, ca.
By
Dave Daniel
·
#147702
·
|
Boonton 92BD Troubleshooting
Boonton stuff ok in this group?? If not, I can delete.
I picked up a mostly dead Boonton 92BD RF Millivoltmeter, seemed like a handy tool should i get it working. I've replaced the electrolytics and
By
Bill Berzinskas
·
#147701
·
|
Re: Anyone able to do GC or GC-MS on Hp ThinkJet Black ink ( for the Hp ThinkJet 2225 printer; Hp 9226a Or Hp 51604a ink cartridges ) ?
Here's one from 2015.? It seems they omitted the Food black two or Black 7984 (a complex diazo dye) which is cataloged by NIH. Not sure if it is the same since it says proprietary and does not give a
By
Frank Mashockie
·
#147700
·
|
Re: Anyone able to do GC or GC-MS on Hp ThinkJet Black ink ( for the Hp ThinkJet 2225 printer; Hp 9226a Or Hp 51604a ink cartridges ) ?
HP MSDS data for 2002, for 51604A Black
Component/Substance CAS Number ?EU Number % by Weight
Ethylene glycol ? 107-21-1 203-473-3 <35%
Triethylene glycol 112-27-6 203-953-2 <15%
Food Black 2 acid
By
Roy Thistle
·
#147699
·
|
Re: Anyone able to do GC or GC-MS on Hp ThinkJet Black ink ( for the Hp ThinkJet 2225 printer; Hp 9226a Or Hp 51604a ink cartridges ) ?
Are you asking for someone to do this for free for you? As a favor?? I don't know how much experience you have with GC-MS (I worked as an analytical chemist for years) but there's a lot of work that
By
Frank Mashockie
·
#147698
·
|
Re: Anyone able to do GC or GC-MS on Hp ThinkJet Black ink ( for the Hp ThinkJet 2225 printer; Hp 9226a Or Hp 51604a ink cartridges ) ?
-- Am? Still?? Looking for anyone able to do GC or GC-MS .... on the Hp black ThinkJet inks:
I am trying to experiment on the inks & the printer for various purposes...
So knowing ? what,? ? &
By
gren
·
#147697
·
|
Looking for POWER keycap for 3586B
Anyone got a parts mule? ?I am looking for the Power button keycap for the 3586B. ?This is the standard "shadow" switch type keycap approximately 0.55 inch x 0.325 inch. ?The old one got lost
By
Tim Wright
·
#147696
·
|
Re: problems with an E4407B spectrum analyzer
Hej Peter
Pr?v og ring til mig
Mvh
Hardy
Fra: [email protected]
<[email protected]> P? vegne af Peter Hansen
Sendt: 4. oktober 2024 11:00
Til:
By
hardyhansendk
·
#147695
·
|
Re: 8131A Error F500
Hi,
Swapping the cards between the two instruments is a good idea, you will have
a better idea which cards are faulty. And if you have a fully functional
generator, you can work on the first one
By
Yves Tardif
·
#147694
·
|
Re: problems with an E4407B spectrum analyzer
Hi,
In my case I did not have the B72 option.
I had bought and installed the missing circuits on the flash card and
repositioned a jumper.
I also used an old PC under XPpro to create the
By
Yves Tardif
·
#147693
·
|
Re: 8131A Error F500
Here's a link I found online to the service manual for 8131A if you don't have it already. ? https://www.doe.carleton.ca/~nagui/labequip/datagen/8131A_Service%20Manual.pdf
By
Frank Mashockie
·
#147692
·
|
Re: E8401A VXI Mainframe Power Supply
That's what I first thought - no signs or smell. I took the filter out and nothing suspicious on the outside.
Thx, Wolfgang
By
Wolfgang Schraml
·
#147691
·
|
Re: problems with an E4407B spectrum analyzer
Hi Peter,
I have used second option Razvan mentioned to upgrade my 4402B (after upgrading flash module by soldering 2 more memory chips).
What I did:
1. download A.14.06 firmware package (.exe)
By
Vladimir Filip
·