8594E CAL: SPAN SENS FAIL
I have an 8594E spectrum analyzer that gives CAL: SPAN SENS FAIL when self-calibrating. I went through the internal service-diagnostic routines and passes:? 2V reference detector, ground reference
By Scott Townley · #143926 ·
Re: Request HDD image for Agilent Infiniium DCA 86100C mainframe
Hi Alex, Thanks again for the image, I'll try it as soon as I get home. What program do you recommend to transfer the image to the HDD? Best regards Maurizio
By maurit · #143925 ·
Re: Request HDD image for Agilent Infiniium DCA 86100C mainframe
Hi Maurizio, I sent you the link. So, and this is my e-mail - aleksandr.tarasov.br@... if someone can send me the 86100C image.
By Alex T · #143924 ·
Re: 54620C not booting
I appreciate the link to the article on the FRAM modification. I have a 54501A headed this direction DOA and once the power supply is sorted I suspect that NVRAM will be an issue. That article saves
By Dave Miller · #143923 ·
Re: 54620C not booting
Thanks for sharing the NVRAMs have been a issue for a long time. Particularly in Tek 2465 scopes. The glitchhawk link seems to do a really great job of showing how to outfit a new battery. At my
By paulswed · #143922 ·
Re: Request HDD image for Agilent Infiniium DCA 86100C mainframe
Hi Alex, I’m very interested in your Win98 HDD image. My 86100A lost some program files. thanks a lot. Maurizio - maurinot@...
By maurit · #143921 ·
Re: 54620C not booting
Hi.. There is plenty of stuff on the interweb, showing how other people managed to remove the battery "Hat" of a NOVRAM, and wired in an external cell. Check these links out... NVRAM to RAMTRON
By Dave_G0WBX · #143920 ·
Re: HP 3585B Spectrum Analyzer How to start diagnosing "No Horizontal sweep" problem
Check your unit. There are apparently two different setups for the horizontal oscillator/HV section. One version has the oscillator and flyback all one one board, which is located under the shielded
By Jeremy B · #143919 ·
HP 5356 B or C frequency converter for 5345A parts needed
I am looking for a 5356B or C parts unit.? Working condition not important, but I'd like the connector and possibly the case to be recoverable.? Don't care about the cable. Anybody have anything
By Bruce · #143918 ·
Re: HP 3585B Spectrum Analyzer How to start diagnosing "No Horizontal sweep" problem
There is a fuse inside the unit on one of the boards. It supplies 12v to the horizontal oscillator. If your flyback is bad, that fuse will blow and you will have no display on the CRT. The board is
By Jeremy B · #143917 ·
Re: HP 3585B Spectrum Analyzer How to start diagnosing "No Horizontal sweep" problem
Do you have an XY monitor? There is video out connectors on the rear of the unit which you can use to obtain video. You can also probe those with an oscilloscope to check for a video waveform if
By Jeremy B · #143916 ·
Request HDD image for Agilent Infiniium DCA 86100C mainframe
Hi guys! Can anyone share the disk image for the 86100C? This mainframe runs on Windows XP. I have a disk image for the 86100A based on Windows 98 and I'm ready to share if anyone needs it. Alex.
By Alex T · #143915 ·
Re: Requesting HP 33120A VFD
yeah I already found some on eBay for the entire front panel assembly. I was hoping someone had just the VFD to spend a little less money. (Believe or not but I had already change the front panel
By alberto.vaudagna · #143914 ·
Re: Requesting HP 33120A VFD
Keysight lists this part as obsolete and cannot be purchased. Keep a lookout on eBay and pick up a front display assembly or low cost parts unit (hard to do these days). [rant on] Keysight has
By Ctron5890 · #143913 ·
Re: 5342A: sometimes counts, sometimes not
Acording to service manual, divide-by-N circuitry are ok. Found faulty component: U2 in A10, a phase detector (Motorola MC4044P). It has two inputs, divide by N signal which is present and changing
By RubenRubio · #143912 ·
Re: 8711B tantalum disease
One was a 20 volt part on a 15 volt rail.? The other was a 10 volt part (!) on a 12 volt rail. I replaced both with 35 volt parts. Nice that the power supplies have current sensing and shut
By Peter Gottlieb · #143911 ·
Re: 8711B tantalum disease
My impression is that the ones that fail are those that have voltage applied to them which is close to or at the part's rated voltage. One instrument that I'm familiar with used a *lot* of parts
By Roy J. Tellason, Sr. · #143910 ·
HP3336C micro controller board
Is board 03326-66516 a replacement/upgrade from board 0336-66505? Thanks -- Joe White KW4YW
By Joe White · #143909 ·
Re: Agilent E4404B Spectrum Analyzer With Dead PSU Repair
Update, the unit powers on. But now is showing other errors. I will create a new thread
By Nick Stanley · #143908 ·
Re: 5342A: sometimes counts, sometimes not
Following steps on service decission tree: it seems to have proper signal forms and values. As you see in the photos of my scope, signals described in the service captures attached are present, but no
By RubenRubio · #143907 ·