CHECK OUT THE WIKI The purpose of the wiki is mainly to allow you to find information on instruments, either from either
- The model number
- The function(s) listed below. Some instruments have multiple functions - for example, the 4195A is a VNA, spectrum analyzer and an impedance analyzer. Therefore the 4195A is listed in multiple categories
Please also check out HPWiki available here:
- Accessory kits - various types
- AC power analyzers - PA2201A and PA2203A
- AC power supplies 6811C, 6812C, 6813C
- Airlines
- Arbitrary waveform generators M8194A
- Amplifiers?493A, 495A?
- Attenuators (optical) 8156A, 8157A, 8158B, 81566A, 81576A,?
- Attenuators (RF) 8494A
- Attenuator set (500 ¦¸) 350C
- Attenuator set (600 ¦¸) 350D
- Attenuator switch driver
- Audio analyzers? 8903A, 8903B, 8903E,? ?
- Base station test sets
- Bit error rate testers (BERTs)
- Cables
- Capacitance meters U1701A, U1701B, 4272A, 4278A, 4279A
- Capacitor Bridge 4270A,
- Capacitor standards 16380A, 16380C,?
- Carrier noise test setsi
- Cesium frequency standards
- Clamp ammeters
- Close field probes
- Crystal Impedance E4915A, E4916A
- Data Acquisition Systems (DAQs)
- DC power analyzers
- DC power supplies 6030A , 6031A , 6032A, 6033A, 6035A, 6131C, 6621A, 6622A, 6623A, 6624A, 6627A, 6255A, 6645A, 6671A, 6672A, 6673A, 6674A, 6675A, 62003A, 62003C, 62003E, 62004A, 62004B, 62004E, 62005A, 62005B, 62005E, 62006A, 62006B, 62006E, 62010A, 62010C, 62010E, 62012A, 62012C, 62012E, 62015A, 62015C, 62015E, 62018A, 62018C, 62018E, 62024A, 62024C, 62024E, 62028A, 62028C, 62028E, 62048A, 62048C, 62048E
- Delay lines
- Detectors
- Device current waveform analyzers
- Digital communications analyzers
- Directional couplers
- Distortion analyzers 330B, 330C, 330D, 331A, 332A, 333A, 334A, 339A, 8903A, 8903B, 8903E,???
- Dynamic measurement DC source
- Electrometers
- Fading simulators
- Femto ammeters
- Filters
- Frequency counters 522B, 5342A 5343A 5352B
- Frequency standards?
- Function Generators ? 3310A,? 8165A,
- GPIB controllers, extenders, cables etc.
- GPS frequency standards
- Harmonic mixers
- High resistance meters 4339B
- High resistance meter fixtures 16008B
- HEV EV Grid Emulators and Test Systems
- In-circuit test systems
- Impedance analyzers 4195A, 4291A, 4291B, 4395A, 4396A, 4396B, 4294A, E4990A, E4991A
- Impedance Analyzer Accessories
- Impedance / Gain Phase analyzer 4194A
- Impedance Meter 4193A,
- Isolators
- LCR meters? U1701A, U1701B, U1731A,? U1731B, U1731C, U1732A, U1732B, U1732C, U1733C, 4191A , 4192A, 4194A, 4195A, E4196A,? 4216A, 4260A, 4261A, 4262A? 4263A, 4263B, 4271B, 4274A, 4275A, 4276A , 4277A, 4284A, 4285A, 4286A, 4287A, 4291A, 4291B, 4294A, 4332A, 4342A, 4395A, 4396A, 4396B, E4980A and E4980AL
- LCR meter calibration devices? 16380A 42030A? 42090A, 42091A and 42100A
- LCR meter accessories
- 2-Terminal BNCs.
- 4-Terminal Pair (BNC connectors)
- Cable extension 16048A, 16048D, 16048E, 16048G, 16048H
- DC current bias accessories 42841A, 42842A, 42842B, 42842C, 42843A
- DC voltage bias accessories 16065A, 16065C,
- Kelvin clips 16089A, 16089B, 16089C,16089E
- Lead Components 16047A,16047B, 16047D, 16047E
- Material 16451B, 16452A
- Probes 42941A
- SMD 16034E, 16034G, 16034H
- 2-port 16096A
- 7 mm (APC7)
- 2-Terminal BNCs.
- LCZ meters? 4276A, 4277A,
- Lightwave clock / data receivers
- Lightwave converter
- Lightwave component analyzer
- Lightwave measurement system mainframes
- Lightwave polarization analyzers 8509B
- Logic analyzers
- Nemo wireless network solutions.
- Noise and interference test set
- Noise figure analyzers
- Noise sources 346A, 346B. 346C ,
- Matching pads (50 ohm to 75 ohm or similar)
- Materials test equipment
- Microwave repeaters
- Microwave downconverters 70427A
- Microwave / THz sources
- Milliammeter 428B
- Milliohm meter
- Mobile communications DC source
- Modular instruments
- AXIe
- Data acquisition (DAQ)
- USB
- PXIe
- Modulation analyzers
- Multimeters 427A, 970A
- Optical attenuators
- Optical heads
- Optical sources
- Optical spectrum analyzers
- Oscilloscopes 120A, 120AR, 120B, 122A, 130A, 130B, 130BR, 130C, 140A, 140B, 141A, 150A, 150AR, 160B, 180A, 180AR, 180CD, 181A, 181AR, 181T, 181TR, 182C, 182T, 183A, 183B, 184A, 184B, 185A, 185B, 1200A, 1200B, 1220A, 1221A, 1703A, 1707A, 1707B, 1710A, 1710B, 1715A, 1722A, 1725A, 1726A, 1740A, 1741A, 1742A, 1743A, 1744A, 1746A, 1980A, 1980B, 5403A, 6000A, 6000L, 16533A, 16534A, 54100A, 5410B, 54100C, 5100D, 54111D, 54120A, 54120B, 54200A, 54501A, 54502A, 54503A, 54504A, 54520A, 54520C, 54540A, 54540C, 54542A, 54542C, 54600B, 54601A, 54601B, 54602B, 54603B,? 54645A, 54654N, 54710A, 54720A, 54750A, 54825N, E1428,?
- Oven controlled crystal oscillators (OCXOs)
- Pattern generators
- PCM terminal test set
- Phase noise measurement
- Pico ammeters
- Printers 2225
- Plotters 7470A, 7475A?
- Probes
- Protocol analyzers and exercisers.
- Power booster test sets
- Power meters 431A, 431B, 431C, 432A, 435A, 435B, 437B, 438A
- Power splitters
- Power supplies
- Pulse generators
- Q-meters 4342A?
- Q-meter calibration inductors 16470A
- Reflection transmission test set
- Return loss module (optical)
- Relays / switches / switch matrices (optical)
- Relays / switches / switch matrices (RF)
- Resistor standards 42030A?and 42100A
- S-parameter test sets
- Scalar network analyzers
- SCSI bus preprocessor interface E2324A
- Selective level meters 3746A
- Semiconductors
- Semiconductor parameter analyzers 4145A, 4155B, 4156B,
- Signal analyzers
- Signal generators / sweep generators / signal sources / oscillators 200CD, 201B, 209A, 204D,? 608A,? 8165A
- Software
- Source measure units
- Spectrum analyzers 4195A,???
- Switch control units
- SWR meter 415E?
- Time interval? counters
- Time mark generator 226A
- Timing and data state modules
- Torque wrenches
- Transmitter testers
- Trigger modules
- Ultrasound transducers
- Universal bridge? 4260A, 4265A, 4265B?
- Vacuum tube voltmeter 410C
- Vector Impedance Meter 4193A, 4800A, 4815A
- Vector Network Analyzers (VNAs) 4195A,? 8510A, 8510B, 8510C, 8753A, 8753B, 8753C, 8753D, 8753E, 8753ES, 8752ET, 8719A, 8719B, 8719C, 8719D, 8720A, 8720B, 8720C, 8720D, 8720ES, 8722A, 8722B, 8722C, 8722D, 8722ES,
- Vector Network Analyzers (VNA) calibration kits 85032B, 85032E, 85033C, 85033D, 85033E, 85050B, 85050C, 85050D, 85052B, 85052C, 85052D, 85054A, 85054B, 85054D, 85056A
- Vector Network Analyzer (VNA) verification kits
- Vector Signal Analyzer 89650S, 89600S
- Vector voltmeters 8405A, 8508A,
- VXI mainframes 70000B, 70000C
- Waveform and function generators
- Waveguide to waveguide and waveguide to coaxial transitions.
- Wireless 58 OTA chambers
- Wireless channel emulators
- Wireless network emulators
- Wireless communication test sets
?
Re: Fixing a sick 4332A LCR meter
I would take either one of you up on that...probably a good idea to have at least one spare on hand in case the one in circuit actually does have a mechanical fault. Sean
By
Sean Turner
·
#104564
·
|
Re: Phase Locking an 8350 Plugin
Hi Don, Thanks for the interest and feedback. I understand the intent but maybe I'm missing something here. Let's say I want to lock at 10 GHz. Wouldn't I need a reference that's 10GHz - 20 MHz? If I
By
Mark Bielman
·
#104563
·
|
Re: Phase Locking an 8350 Plugin
Also 8350B system AUX out is typically +8dBm at the 8350B tune freq. Don Bitters
By
Don Bitters
·
#104562
·
|
Re: Phase Locking an 8350 Plugin
The intent is to get the stability of the 8640B applied to the 8350B system. This is done by locking the 8350B to a harmonic of the 8640B freq. ( ie. the 8640B freq. is 1000MHz @ +13dBm then the 8350B
By
Don Bitters
·
#104561
·
|
Re: HP 3455a DVM repair
My guess is that Q12 is shorted Collector to Base and CR15 is what is pulling TP4 down to -18V and caused R14 to overheat. Q12 is listed as an HP 1854-0087 (75MHz ?360mW NPN).
By
Ed Giaimo <edgiaimo@...>
·
#104560
·
|
Re: 8510C questions and cabling
OK - You win, but the daisy chain cables are expensive. I'b be tempted to get a long 8502 - test set cable so you could "remote" the test set, which would make exchanging test sets easier. If you
By
Bruce
·
#104559
·
|
Re: Fixing a sick 4332A LCR meter
If you need that FET, I can send you one... Dan in Chandler, AZ
By
Daniel Nelson
·
#104558
·
|
Re: Fixing a sick 4332A LCR meter
Stephen, 1855-0082, TRANSISTOR:FIELD EFFECT P-CHANNEL Sean
By
Sean Turner
·
#104557
·
|
Re: Fixing a sick 4332A LCR meter
Seam, Please send the p/n I¡¯ll check stock Regards, Stephen Hanselman Datagate Systems, LLC
By
Stephen Hanselman
·
#104556
·
|
Re: 8510C questions and cabling
Well, to have an S parameter set, a TR set and an ISM band antenna range. Do I really need all that? No. But what else am I going to do? Watch TV? Not something that appeals to me. Have Fun! Reg
By
Reginald Beardsley
·
#104555
·
|
Re: 8510C questions and cabling
I found an 08513-60004 on ebay. Has anyone done an opt 001 upgrade on a non-opt 001 8511A, 8512A or 8515A? It would appear to be just recabling and adding the multiplexer boards. Of course, I'd also
By
Reginald Beardsley
·
#104554
·
|
Re: 8510C questions and cabling
You might consider why you actually need all three test sets. Multiple test sets does increase complexity and causes losses in the signal from the sig gen. Cheers! Quoting "Reginald Beardsley via
By
Bruce
·
#104553
·
|
Re: 8510C questions and cabling
The "Preflight Checkout" diagram is clear about where things go, but it presents every variant on the same diagram with cryptic notes. so deciphering the connections for a specific set of modules is
By
Reginald Beardsley
·
#104552
·
|
Re: 58503A GPSDO PSU details/availability
Excellent info thanks Jeremy! Yes please send wiring details. I'm hopefully a few weeks away from completing my shack, and freq references are 1st on the list for loading into the rack. :^)
By
Andy ZL3AG
·
#104551
·
|
Re: Fixing a sick 4332A LCR meter
Hi Chuck, Good point. So far it's stable. I aligned it as best I am able for now, and it's working in all modes with reasonable accuracy. If it becomes intermittent, looks like it is still possible to
By
Sean Turner
·
#104550
·
|
Re: Fixing a sick 4332A LCR meter
Just a repair note: I have seen a lot of TO92 packaged transistors where one or more lead has become intermittent. I believe this is caused by fiddle fingered folks that just have to push the
By
Chuck Harris
·
#104549
·
|
Re: Fixing a sick 4332A LCR meter
Joel, Thanks for the detailed steps on that. I believe I found the smoking gun, a cold solder joint. I removed A1Q1 to test it out of circuit, which really didn't lead to anything. But it started
By
Sean Turner
·
#104548
·
|
Re: Fixing a sick 4332A LCR meter
Sean, I think we're making progress ! In Capacitance mode, the oscillator should run at 100 kHz in the left-hand side of the range switch, and at 1kHz on the right-hand ranges. The frequency switch is
By
Joel Setton
·
#104547
·
|
Re: 8510C questions and cabling
You can daisy chain the test sets IF the test sets have the proper internal connections. Look to see if both "D" connectors have coaxial fittings, if so you can daisy chain. Test sets with this option
By
Bruce
·
#104546
·
|
Re: 8510C questions and cabling
The statement about the BNC open is an awkwardly worded statement, likely by an English challenged individual. All they are trying to say is don't connect anything to that BNC connector... especially
By
Chuck Harris
·
#104545
·
|