Microwave Mixer
jfowler306
Hi all. A bit off the HP theme, however need some help. I have a Time
63132 rf (active) mixer. It also has an alt. part number of DXM602. The LO and RF appear to be in the 10-20GHz range. If anyone has, or can direct me to where I can find specs. on this unit, I would really appreciate it. Eddy W5WTN/K5IPA |
Re: deskjet 895 cxi printer
Don Black
One stock fault in HP printers is the print head slide bar becoming sticky. That's the chrome rod about 1/4" dia that running across the printer. Try wiping it clean with a tissue, that usually clears the fault. It shows up as all kinds of errors.
toggle quoted message
Show quoted text
Don Black. Jerry Massengale wrote: Greeting, |
Re: Workbench surface
At 10:30 AM 6/12/06, you wrote:
I didn't realize that fluorescent lighting could be a problem. In anyI forgot to mention that it was only a real problem because it was quite low over the bench and the cables to the lights wound around the back and side of the bench, plus they were the old inductor type ballast and the large metal housing might not have been well grounded. When a 20v max mosfet gate regularly gets 27v at 50hz it doesnt last long but all other ICs were ok, a drop in the mat resistor to 100k didnt upset any other handling and dropped the field voltage from 27 to around 2.5 I will take a look at 3M and other mats, both for the floor and theAh - I just checked my post, the '3m' later on referred to a 3 metre area of transition (miniumu) from an unsafe to the fully safe area. For me this did not need a 3M mat just an ordinary wool/cotton carpet which was treated with the same chemical women use when ironing to stop their dresses sticking to their panty hoses - it was *much* cheaper than the 3M or any other chemicals, I actually still have a Bokhara Afghan rug, cost about $600 in the olde days when these were being sold as "investments", still in one piece must have had it for over 20years now, more utilitarian than worth hanging on the wall for a seance ;) From what I recall, the chemical is just some ionic salt, perhaps even MgSO4 or maybe some mix which was very slightly hygroscopic, or at least had some lower resistance when dry at low humidity levels cant recall exactly. I dont use it anymore as my smallest lab is for my use only and if I ever have any very badly static sensitive devices on that 3M bench then my whole demeanor changes dramatically :P Cheers Mike 73, |
Re: Workbench surface
Steve
Thanks Mike and others for the advice, it's most helpful and educational.
toggle quoted message
Show quoted text
I didn't realize that fluorescent lighting could be a problem. In any event, I'm not using that kind of lighting. I will take a look at 3M and other mats, both for the floor and the bench top. 73, Steve K8JQ Mike wrote: have been through this a couple of times in last 20 years, |
Re: deskjet 895 cxi printer
John Miles
Good site for that type of issue is . They
toggle quoted message
Show quoted text
sell service kits for most HP and other models, at least for Laserjets. Pickup rollers seem to be a common trouble spot. -- john, KE5FX -----Original Message----- |
Re: HP mms system catalog
Hi Erik,
You can find a copy of "Modular Measurement System HP7000 Family Catalog" at the TestMart website ( ). At the bottom of the page, you will see a section called: "Library". In this section. click on the link: "Original manufacturers' catalogs" Now, in the Agilent catalog listings, about 1/3 down the page, you will see the "Modular Measurement System HP7000 Family Catalog" link, where you can download the document. The website has a registration procedure, but it is very easy. Good luck! Matt |
Re: HP1652B Logic Analyser ET AL - system disk, how to create?
MKawasaki
Alexander,
This is another product that I used to work on during my days in Colorado Springs :-) I can't find the file within our intranet, but will send a request to the division support engineers. I plan to get an update within the next week. Mike --- In hp_agilent_equipment@..., Alexander Whiplash <alexander_whiplash@...> wrote: _____________________________________________________________________ _______________ Do you Yahoo!? |
Re: Workbench surface
Rasputin Novgorod
The only solution is to follow a good protocol. I put that in placeAfter several careless incidents with soldering irons, I started using fireproof table tops (with static mats on top). It was some kind of laboratory table-top, a mix of asbestos and resin, that was made locally when I lived in New England. I've had them for 20 years. I remember the factory cutting room was clouded with airborne asbestos and cement dust; nobody wearing masks. The safety inspectors wound have had fits... /b ____________________________________________________________________________________ Have a burning question? Go to www.Answers.yahoo.com and get answers from real people who know. |
Re: Workbench surface
have been through this a couple of times in last 20 years,That was all very interesting - particularly that a 1M-ohm mat was susceptible to fluorescent light e-m. Two points that are worth making to clarify: 1. The reason that Al foil is bad is that any static charge on a device comes out instantly - and the high peak current that flows (even from a few pF on the gate of a FET) is so high that thermal damage occurs. A resistive mat is essential to make sure that charge bleeds out slowly enough to prevent this. 2. It is possible to subtly damage devices without destroying them totally - leading to change in characteristics, noise and/or lifetime. The only solution is to follow a good protocol. I put that in place after I destroyed a rare double FET while repairing a 'scope. Craig |
Re: Workbench surface
have been through this a couple of times in last 20 years,
toggle quoted message
Show quoted text
a. Using a static mat from 3M or one of the top companies that supply to HP and others is the best recommendation, those mats allow static discharge at the best rate. Dont use an aluminium sheet to ground as some people have tried. b. If you have overhead fluoros then might need to reduce the mats 1M to 100K, in my case the 1M meant the mat wasnt draining the field from the fluoro and we blew many mosfet gates (the ones without protection zeners). If you are not going to be handling anything without protection zeners then dont bother. c. Have a region of interface between the static mat region and where you walk from so some static can dissipate before you get to the mat. Like a 3m area which is floor treated, floor mat is ideal but I find that carpet is ok if its treated twice a year with some spray on static treatment. d. If you are handling a lot of very sensitive devices in most other places in the lab where they might be stored but not worked on then a static controlled fan might be helpful, cant recall the name but it has a ion source either mildly radioactive or a power operated type. I dont use one myself but depends what you are handling, others might comment on that. cheers mike At 11:50 AM 3/12/06, you wrote:
There was a discussion regarding workbenches on this reflector back in |
Freebies Update
All computer items (the Proliant and the Suns) are now spoken for.
The plotter is still available. Thanks much. -=-=-=-=-=-=-=-=-=-=-=- Bruce Lane, Owner & Head Hardware Heavy, Blue Feather Technologies -- kyrrin (at) bluefeathertech do/t c=o=m "If Salvador Dali had owned a computer, would it have been equipped with surreal ports?" |
FREE Equipment in WA State (Kent)
Fellow Techies,
I've said it on my web site, and I'll say it here: Scrounging is not so much ongoing acquisition as it is a 'flow-thru' of goodies. Any tech-scrounger will eventually hit the point where they Just Need To Move Stuff Out. Such is the case with me right now, so I'd like to take a moment here to feed back to the same 'techie food chain' that I've benefited from for so many moons. Specifically: I've got the following FREEBIES available for PICKUP ONLY, NO SHIPPING in Kent, Washington (southeast of Seattle). The only condition I place on the stuff is that whoever takes it needs to be taking it to USE IT, not to sell it for scrap. --HP DraftMaster SX Plus large-format pen plotter, in excellent condition. Includes a whole slew of pens (condition unknown, but they've been kept capped in an indoor environment) and two packages of big paper (including a pack of Size D sheets). This item is not that heavy, and it is on wheels, but it is bulky and tall. You will need at least a small pickup truck with an open bed to move it properly. --Compaq ProLiant 6500 server with twin Pentium Pro 200's. This was the box that was originally running my FTP archive (I've since upgraded to a later model of ProLiant). Working condition, 512MB RAM, and I think I could find a pair of rack slides and maybe even enough 4 gig drives to make a small array if you twist my arm enough. --A pair of Sun SPARC-10 systems with at least 50MHz HyperSPARC CPU modules installed. These used to run part of my Internet presence before I upgraded to HP NetServers, so I know they work. RAM and hard drives included (though I will need to wipe the drives before they're taken as they still contain security-sensitive data). If you'd like any or all of the goodies (hopefully all in one swell foop?), just drop me a line. If, for whatever reason, your E-mail bounces you can give me a buzz on 253-639-2996. Thanks in advance for helping me find this stuff a good home. -=-=-=-=-=-=-=-=-=-=-=- Bruce Lane, Owner & Head Hardware Heavy, Blue Feather Technologies -- kyrrin (at) bluefeathertech do/t c=o=m "If Salvador Dali had owned a computer, would it have been equipped with surreal ports?" |
Re: HP1652B Logic Analyser - system disk, how to create?
...And a mirror of these files will soon be available on the Blue Feather FTP archive.
Keep the peace(es). * REPLY SEPARATOR * On 27-Nov-06 at 20:38 Warren Guinn wrote: Andreas, -=-=-=-=-=-=-=-=-=-=-=- Bruce Lane, Owner & Head Hardware Heavy, Blue Feather Technologies -- kyrrin (at) bluefeathertech do/t c=o=m "If Salvador Dali had owned a computer, would it have been equipped with surreal ports?" |
Workbench surface
Steve
There was a discussion regarding workbenches on this reflector back in October and I'd like to ask for some follow up advice.
I'm planning to replace an old workbench. Is there a top I can put on my new workbench that is ESD safe? Right now I have an ESD mat that I place on the workbench surface when appropriate. Inasmuch as I'm going to be more or less starting from scratch, I thought I'd at least ask the question about an ESD-safe work surface. Thoughts or advice out there? Thanks. Steve K8JQ |
Re: OT fluke 8060A battery eliminator
Steve
Thanks for the responses and advice guys.
toggle quoted message
Show quoted text
Upon closer inspection, the writing on the meter does say A81. Thanks for that clarification. I thought that isolation might be an issue with the run-of-the-mill wall warts. I'll most likely use an on-hand wall wart for lower voltage measurements, which is what I use the meter for mostly when it's on my workbench. Over the past few years I've often forgotten to turn it off and end up using enough 9V batteries to make it worthwhile to consider powering it from AC. Wonder what size or type of plug fits the battery eliminator jack on the meter? Thanks again! Steve K8JQ jbarnes wrote: Steve, |
Re: Need logging help with hp 5328a
Jeff Spence
James,
toggle quoted message
Show quoted text
The TCL scripting language has good support for GPIB. It is pretty easy to write a script in TCL that reads the 5328A once every so often and writes the result to a text file. The text file can then be read by Excel for further processing. Jeff -----Original Message-----
From: James Hickox [mailto:jameshickox@...] Sent: Friday, December 01, 2006 11:37 AM To: hp_agilent_equipment@... Subject: [hp_agilent_equipment] Need logging help with hp 5328a Hello all, I have a 5328A counter with the HP-1B option 011 digital I/O init installed. I wish to record the frequency output of a oscillator over a long period of time (several hours to several days) and I need some way to interface the output from the counter to a PC, and some way to copy the input to the pc to an excel spreadsheet for data logging, etc. BTW, I am NOT an engineer, just a ham radio operator trying to get a stable ransmitter on the air. Hny Ideas, suggestions would be a great help. Thanks, James Hickox, AA5AO -- No virus found in this incoming message. Checked by AVG Free Edition. Version: 7.5.430 / Virus Database: 268.15.3/562 - Release Date: 12/1/2006 1:12 PM -- No virus found in this outgoing message. Checked by AVG Free Edition. Version: 7.5.430 / Virus Database: 268.15.4/563 - Release Date: 12/2/2006 9:59 AM |
Need logging help with hp 5328a
Hi James:
Spark Fun has the GPIB-USB Controller. In addition to the EZ GPIB software there's also a "Plotter" program at: that's designed to process time interval data (like you would measure with the HP 5328) and from that get Allan plots. Note for high stability work you don't have time to make frequency measurements and instead make Time Interval measurements, typically comparing a GPS timing receiver's 1 PPS with a 1 PPS derived from you oscillator. You might also consider joining the Time Nuts mailing list Have Fun, Brooke Clarke, N6GCE -- w/Java w/o Java |