Re: Crystal detector with 10dB attenuator in series - is it for matching purposes?
Good paper on the subject. https://www.minicircuits.com/app/AN70-001.pdf Greg
By Greg Muir · #146525 ·
Re: 4276A LCZ analyzer needs cal/adjustment,paper manual, standards.
Might Not be relevant, ... since applies to the 4274a,? but: Just noticed an Hp Service Note in the Hp 4274a LCR manual (4274a-12) : " Improvement for Memory Backup Battery " -- about swapping out
By gren · #146524 ·
Re: 8566B Calibration
Okay, got it. Many thanks.
By Jinxie · #146523 ·
Re: 8566B Calibration
=== It is the adjustment trimpot just to the left of RF input. Between LO out and RF input. Ozan
By Ozan · #146522 ·
Re: 8566B Calibration
AMPTD CAL is one of the trim pots on the front panel. The other is FREQ or FREQ ZERO. I cannot remember exactly. Look through the other Recall x instructions and one will be for the frequency pot.
By billyatams · #146521 ·
Re: Crystal detector with 10dB attenuator in series - is it for matching purposes?
The 10dB attenuator is probably recommended as it will help to improve the VSWR of the source impedance of the sig gen that is used to test the 8473C detector. This will help to minimise the overall
By jmr · #146520 ·
Re: 8566B Calibration
In step 3, it says "adjust". Look around the bottom box front panel and you will find it Vladan
By pianovt · #146519 ·
Hp Schottky diode HP QSCH 1691 ( or HPQSCH-1691 ?) ... Data Sheet wanted
Can't find any info on this Hp QSCH 1691 ( Schottky ) diode. It does not seem to be mentioned in either of the 1972, 1980, nor 1984 Hp Diode & Transistor Designer's Handbooks. Anyone have a .pdf
By gren · #146518 ·
8566B Calibration
Hi all, The calibration procedure for this analyzer states that something called "AMPDT CAL" requires adjustment. However, I can find no such button on my particular analyzer. Everything else the
By Jinxie · #146517 ·
Re: Crystal detector with 10dB attenuator in series - is it for matching purposes?
Saevar, One reason for the attenuator may be to provide a DC path to ground.? Many detectors require this, and some RF sources are AC coupled. --John Gord
By John Gord · #146516 ·
Re: HP 141T Power Supply Help Needed
Bill, It sounds like you are on the right track! So find out were the .04Ohm wire goes¡­ Mike Sent: Saturday, August 31, 2024 2:29 PM To: [email protected] Subject:
By Michael Bafaro · #146515 ·
Re: HP 141T Power Supply Help Needed
Neat trick checking for shorts. As you expected probably, 0.04ohm. I have found one purple/white wire which when removed brings that up to 384 ohm. I'll be out of pocket for a day but will look
By Bill Berzinskas · #146514 ·
Crystal detector with 10dB attenuator in series - is it for matching purposes?
I was reading about the crystal detector HP 8473C and in the performance test section in the documentation, it is recommended to use 10dB isolation attenuator in series with the RF port and adjust the
By saevar · #146513 ·
Re: HP 5086-7906 YTO
Yves, thank you very much for all the suggestions and information with which I will continue my investigation In the meantime I received the 2.5mm socket wrench and after a little mod to adapt it for
By alfa beta · #146512 ·
Re: HP 141T Power Supply Help Needed
Bill, So you replaced the pass transistor and Q9 then it popped the 3A fuse F4. And at that point the regulator worked but there must be a short on the -12.6V buss. Please run an Ohm meter check
By Michael Bafaro · #146511 ·
Re: 16901A 16901-14102 recovery disc ISO wanted (was Re: [HP-Agilent-Keysight-equipment] Wanted: 16902B recovery DVD, p/n 16902-14100)
Maybe only on certain interfaces? There's definitely nothing wrong w their configuration when you're close (network and latency-wise) to them. I mean, they often saturate the SSD disk writing
By Keith Monahan · #146510 ·
Re: 16901A 16901-14102 recovery disc ISO wanted (was Re: [HP-Agilent-Keysight-equipment] Wanted: 16902B recovery DVD, p/n 16902-14100)
I wonder whether internet archive simply has TCP 1323 window scaling options disabled or clamped to a low maximum, or something like path mtu discovery broken by blocking ICMP type 3 code 4 packet too
By Ed Marciniak · #146509 ·
Re: 16901A 16901-14102 recovery disc ISO wanted (was Re: [HP-Agilent-Keysight-equipment] Wanted: 16902B recovery DVD, p/n 16902-14100)
Here's the images I have https://www.dropbox.com/scl/fo/05b2wr6le685c1dacwnw9/AMi9MTuEqv4IDBUyAQY2004?rlkey=4e1yka1xbnkhx6iiclziecntq&st=37jkmd83&dl=0 File sizes and MD5sums -rwxr-xr-x 1 root root
By Keith Monahan · #146508 ·
Re: HP 141T Power Supply Help Needed
Those numbers just don't make any sense. I pulled all the transistors AGAIN, desoldered and cleaned. Tested them all with a meter, repopulate. Back to 23.5v.... [email protected]> wrote:
By Bill Berzinskas · #146507 ·
Re: HP 5086-7906 YTO
I forgot to send the mixer information. Blue wire, PIN SW: band0= +9.81v and band 1=-9.56V Green wire, BIAS: +3.9V Yves De : [email protected]
By Yves Tardif · #146506 ·