Re: Agilent E4404B Spectrum Analyzer With Dead PSU Repair
Assuming normal operation for a power supply, either overvoltage or overcurrent conditions.? If you have a scope, look at the rail voltage vs the shutdown signal and see what's going on and when.
By Harvey White · #143856 ·
Re: Agilent E4404B Spectrum Analyzer With Dead PSU Repair
My previous idea about why the controller circuit shutdown is wrong. I measured the IN-OUTPUT of the LT1121. The input changes between 13-20V, which is more than enough headroom for a 10V output
By Nick Stanley · #143855 ·
Re: HP6643A - Blows Fuse When Activated
That is a 210W power supply. You need at least 200W of lamps. [email protected]> wrote:
By Michael A. Terrell · #143854 ·
Re: Agilent E4404B Spectrum Analyzer With Dead PSU Repair
Hi Eric, I believe it's this cap (25V - 470uF). It's connected directly over the supply line of the LM339 and LT1121
By Nick Stanley · #143853 ·
Re: HP6643A - Blows Fuse When Activated
First, you need to chose a bulb that will draw ALMOST as much current as the fuse is rated for. That way, if there's a short in the DUT (device under test), all the voltage drop will fall across the
By Jim Adney · #143852 ·
HP Linefinder
Good Morning, Does anyone have any information about the HP 11404A Linefinder unit. I believe it is used with the HP surveying equipment (3805, 3820 etc) but I have no idea how. It looks like it would
By Gordon Batey · #143851 ·
Re: N9020A
Hi Just forgot To get into service mode use -2061 Hardy Fra: [email protected] <[email protected]> P? vegne af jmr via groups.io Sendt: 25. april 2024
By hardyhansendk · #143850 ·
Re: N9020A
Hi John You can print several screen types to usb stick. .or internal. Hardy Fra: [email protected] <[email protected]> P? vegne af rftek.ltd via groups.io
By hardyhansendk · #143849 ·
Re: N9020A
Hi Jeremy It do have inbuilt dc block,and now john have the admin keys to use. Its straight forward to see the lic. files¡ªalso recommend to use the inbuilt usb based backup program.And as john
By hardyhansendk · #143848 ·
Re: Agilent E4404B Spectrum Analyzer With Dead PSU Repair
Hello Nick, What is the startup cap ? Is it the group of two 680?F capacitors C189 & C190 ? Best regards Eric
By f1ghb · #143847 ·
Re: N9020A
John Great We have a talk off list¡­. The service manual clear says To get into the system use 1¡­administrator 2¡­agilent4u Then you have all rights Regarding vsa 89600¡­off list. Make a usb
By hardyhansendk · #143846 ·
Re: Agilent E4404B Spectrum Analyzer With Dead PSU Repair
Welcome to the club. Even if you have removed all the RIFA caps now, there are still electrolytics of unknown status, so here is my PSA for the day: *** Do wear eye protection when working on power
By Carsten Bormann · #143845 ·
Re: Agilent E4404B Spectrum Analyzer With Dead PSU Repair
Apologize for the 2MB size photos, I was in a rush. And also, the SMPS was powered from an isolation transformer
By Nick Stanley · #143844 ·
Re: Agilent E4404B Spectrum Analyzer With Dead PSU Repair
After a bit of reading I sort of understand the start up procedure of a typical SMPS. A start up capacitor is charged from rectified mains, usually via a power resistor. This cap is connected to the
By Nick Stanley · #143843 ·
Re: Welcome, Leo! I know I've seen your name before; are you on the TekScopes group? That's a nice HPAK gear collection you have. Unfortunately I can't help you with your 54620C. I had a 54504A for a while and currently have a 54120B. All my other scopes (4 at the moment) are Tektronix. I fix want to welcome you to the HPAK group, though. Jim Ford, Laguna Hills, California, USA Sent from my T-Mobile 4G LTE Deviceber
I am. :)
By Leo Potjewijd · #143842 ·
Re: N9020A
Sounds good! It would be interesting to see a screenshot of the hardware page where it lists the hardware modules inside the analyser. Somewhere on the HDD there will be all the *.LIC files for your
By jmr · #143841 ·
Re: N9020A
Well mine has arrived and the reason the whole tread was started, which hasn't been hijacked too much but would like to report on my findings on the unit delivered. Seems OK, bit scruffy but fully
By rftek.ltd@... · #143840 ·
Re: N9020A
In the case of Alltest, it's normal for it to have both names, since you can see that the processor board has been changed, since the instrument is now running Windows 7. At the time of its release,
By F1EKU · #143839 ·
Re: HP6643A - Blows Fuse When Activated
You may need a higher wattage bulb ¡­ What is the rated current draw of the DUT? I have a whole set of bulbs up to 200W and can parallel them for more if necessary. D. Sent: 25 April 2024 03:02 To:
By David C. Partridge · #143838 ·
Re: N9020A
I have no financially interest in this-but please look up the seller I refer to.He clearly states that their MXA`s have been ``fixed`` regarding options,so no Keysight Lab will service/cal it..The
By hardyhansendk · #143837 ·