Re: 8595E YTF problem (loss high band)
Please do not remove the mesh. Filter will be damaged without. There will come alot of spurious if the mesh cannot be mounted again. br Peter ________________________________ Fra:
By
Peter Hansen
·
#140727
·
|
Re: 8595E YTF problem (loss high band)
Be extremely carefull. most likey you will ruin your YTF totally. Adjusting a YTF is a very hard job. I do it on the different series of analyzer. But it requires a nwa and some special tools. And it
By
Peter Hansen
·
#140726
·
|
Re: 8595E YTF problem (loss high band)
To answer your question: I have had this Spectrum for several years, and on occasion I do some performance tests, and in particular, a flatness test. In 2021, I removed this Spectrum from my test
By
Yves Tardif
·
#140725
·
|
Re: HP 8920A Cal and CRT replacement
Brilliant display!
By
RFI-EMI-GUY
·
#140724
·
|
Re: 8595E YTF problem (loss high band)
Don't rush to judgement too fast. The YIG spheres appear to still be mounted on-axis with the sapphire mounting rods, which are typically adjustable in some way. Both the insertion depth and rotation
By
Ed Breya
·
#140723
·
|
Re: Help with HP 3335A output jitter
I hope you can resolve what is wrong with your 3335A. I just serviced mine due to a failed internal frequency reference (10544B). Anything out side of that and they are a PITA to service correctly. I
By
Craig Petersen
·
#140722
·
|
Re: HP 8920A Cal and CRT replacement
Looks very nice ! Congrats!! I¡¯ve one too Ing. Patricio A. Greco Taller Aeron¨¢utico de Reparaci¨®n 1B-349 Organizaci¨®n de Mantenimiento Aeron¨¢utico de la Defensa OMAD-001 Gral. Mart¨ªn Rodr¨ªguez
By
Patricio A. Greco
·
#140721
·
|
Re: 8595E YTF problem (loss high band)
This confirms what I feared, thank you for your comment. Yves De : [email protected] <[email protected]> De la part de amirb Envoy¨¦ : 11 d¨¦cembre 2023
By
Yves Tardif
·
#140720
·
|
Re: HP 8920A Cal and CRT replacement
Well, it took a couple of years but I finally pulled the trigger and installed NewScope LCD CRT retrofit.? The instructions where spot on, the results are fantastic.
By
Damon - KJ7E
·
#140719
·
|
Re: 8595E YTF problem (loss high band)
the spheres have moved and the way they are positioned has weakened the coupling between the adjacent cavities that explains the excess IL 15dB excess IL cannot be compensated by alignment.
By
amirb
·
#140718
·
|
Re: HP 11567A 20 CM Air Line - looking for datasheet
Yes, I saw that. On two instruments I have, I measure 4.96 degrees at 10 MHz (for a short circuit phase shift (factor of 2)). By my calculation that (2.48 deg) equates to 20.65 cm (at a Er of 1.0) ..
By
vk2bea
·
#140717
·
|
Re: HP 11567A 20 CM Air Line - looking for datasheet
Hi Micheal, From Hp catalog 1976: I found this, I don't know if it can help you. Regards Yves De : [email protected] <[email protected]> De la part de
By
Yves Tardif
·
#140716
·
|
HP 11567A 20 CM Air Line - looking for datasheet
Does anyone have data on the?HP 11567A 20 CM Air Line (APC-7) beyond the information in the HP catalog? What is the tolerance on the nominal length (20.21 cm)? Can one assume a dielectric constant of
By
vk2bea
·
#140715
·
|
8595E YTF problem (loss high band)
Hi, I have a problem with my 8595E, the 2.9-6.5 GHz band has too much insertion loss, around 15dB flat. The low band works correctly. I know that this seems to be a common problem, and the question
By
Yves Tardif
·
#140714
·
|
Parting out the remaining parts of an 8558 Spectrum Analyzer
Howdy folks, Some time ago, in this forum, I described how I repaired the contact fingers of an 8558 spectrum analyzer without taking the front panel off (the involvement and commitment of taking that
By
Daniel Koller
·
#140713
·
|
Re: HP 5344S Microwave Source Synchronizer
I cannot read what the span is on the SA, is it 2MHz, or 2KHz? If 2 KHz span, then it looks like you have 60Hz noise or spurs. For this case check or ground all your chassis together, operate them
By
Don Bitters
·
#140712
·
|
HP 5344S Microwave Source Synchronizer
Hello everyone, I have used the HP 5344S Microwave Source Synchronizer consisting of 5343A and 5344A on the HP 8350B. So far the setup works well. The 5344S locks the signal and I get a very accurate
By
Rudolf BoscoLab
·
#140711
·
|
Re: HPWiki
Always wondered why a HPAKwiki (in the vein of TekWiki) wasnt started earlier, so good work Dave!
By
Nik Milosevic
·
#140710
·
|
Re: "Read the documentation"
You have a fundamental misunderstanding of how search engines work. People using the site do not affect search results. People searching for it and then going to it in search engine results affect
By
Dave McGuire
·
#140709
·
|
"Read the documentation"
Argentina has the world's highest official rate of inflation. Zimbabwe changed how they calculated theirs to put Argentina in the top spot. I thought I'd hire some hams in Argentina at low wages to
By
Reginald Beardsley
·
#140708
·
|