¿ªÔÆÌåÓý

Date
Re: Problem with HP MMS 70000, USERLOCK ON
And I measured the effective retention voltage on the vdd of the TC551001CF-70L memories, it's at 2.864v , the datasheet says it should be at 2v minimum to keep up the data, so definitively it's not a
By Pierre-Fran?ois (f5bqp_pfm) · #105400 ·
Re: Problem with HP MMS 70000, USERLOCK ON
I then decided to open my 70900B, have not done this since many years when I upgraded the firmware with the latest from my previous 70952B. At that time I uploaded this latest firmware on KO4BB web
By Pierre-Fran?ois (f5bqp_pfm) · #105399 ·
Re: HP8562A Errors 304,335,317,333
Sound advice. Thank you very much. Stu From: [email protected] <[email protected]> On Behalf Of Peter Hansen Sent: 22 April 2020 08:47 To:
By stur_mm · #105398 ·
Re: HP8562A Errors 304,335,317,333
Hello Stu be very carefull if you will recap the psu. It is an old brittle 4 layer board. solder joints does need alot of heat. You can easy ruin the connection to the middle layer. The 3 legged cap
By Peter Hansen · #105397 ·
Re: Series 1 & 2 boxes (was Seeking Spring Nuts for Series 1 cabinet)
The System II designation refers to the second design of MODULAR case. Not all HPAK instruments use modular cases. Obviously the ones pre-dating? it? can't. 339A and 3456A both use modular cases,
By Robert G8RPI · #105396 ·
Re: HP8562A Errors 304,335,317,333
Steve, Waiting for some ¡°Mate N lock¡± plugs to Test an existing board I have funny config but supplies all relevant circuits (RF) associated with errors I¡¯m seeing. Any ideas on the 3 legged
By stur_mm · #105395 ·
New file uploaded to [email protected]
Hello, This email message is a notification to let you know that the following files have been uploaded to the Files area of the [email protected] group. * /3000 to 3999/HP 3314A
By [email protected] Notification <HP-Agilent-Keysight-equipment+notification@...> · #105394 ·
Re: Series 1 & 2 boxes (was Seeking Spring Nuts for Series 1 cabinet)
Some photos of System I and System II Cases on my website.HP Cases | | | | HP Cases HP Cases for test equipment System I System II Spare parts for sale ? Azur Electronics HP Cases for test equipment
By Steve A · #105393 ·
Re: new File called App notes
david.partridge@...> wrote: At least some of it is a result of taking active design steps to hinder counterfeiting. Dave. -- Dr. David Kirkby, Kirkby Microwave Ltd,
By Dr. David Kirkby, Kirkby Microwave Ltd · #105392 ·
Re: Problem with HP MMS 70000, USERLOCK ON
Hi JF & the Gang, After reloading a fresh DLP NF 70875A from my original ROM personality module I got the same error ERR 2001 and entered once again into the USERLOCK ON dead lock. I was again obliged
By Pierre-Fran?ois (f5bqp_pfm) · #105391 ·
Re: Decline, was Re: [HP-Agilent-Keysight-equipment] new File called App notes
Lack of understanding of low level hardware is surprisingly common...wrote my latest book because of that. Would, in fact, like to keep this email as a testimonial of sorts... --- With best regards
By Tam Hanna · #105390 ·
WTB: 5265A Digital Voltmeter for HP 5245L Electronic Counter.
Good Morning, Thought I would ask on the chance that a *5265A Digital Voltmeter* plug-in might be for sale here. I plan to install the 5265A Digital Voltmeter plug-in into a *K37-5245L Converter
By Terry Gains · #105389 ·
Re: Maximum depth of HP rack mountable gear?
FWIW, here's my experience and method of cutting a rack down to size. Same problem, it was too deep for the available space. http://everist.org/NobLog/20130923_hd_procrastination.htm When setting up
By Guy Dunphy · #105388 ·
Re: Do ManualsPlus still exist
The loss of ManualsPlus and the other physical archives, was a disaster. Few realize the magnitude. Some links and photos here: http://everist.org/NobLog/20190223_full_spectrum.htm#alex As for
By Guy Dunphy · #105387 ·
Re: Boonton Type 260-A Q Meter
Regarding the 260, mine had a failure in the constant voltage transformer power supply.? While today's power is much more stable than yesterday's, the system also provides line isolation which is
By Bob Albert <bob91343@...> · #105386 ·
Re: Boonton Type 260-A Q Meter
Probably the most common problem with the 260 and the RX meter are dirty contacts in the oscillator. Fortunately they are not hard to get to and clean. A few applications of Deoxit will clean them.
By Richard Knoppow · #105385 ·
Re: HP 3314A manual needed
If possible I¡¯d like a copy, thanks!! S.hanselman@... Regards, Stephen Hanselman Datagate Systems, LLC
By Stephen Hanselman · #105384 ·
Re: HP 3314A manual needed
sorry but the one on keysight has all the chapters including servicing and adjustment and also schematics http://literature.cdn.keysight.com/litweb/pdf/03314-90021.pdf this is the one? I have used to
By amirb · #105383 ·
Re: HP400E was new File called App notes
Oh, yeah. With 50 ohm input termination, that noise probably isn't coming from there. Give those capacitors a stern look. -Dave -- Dave McGuire, AK4HZ New Kensington, PA
By Dave McGuire · #105379 ·
Re: HP400E was new File called App notes
Still unstable with 50 ohm input.? Decoupling caps most likely need replacing.JimSent from my Verizon, Samsung Galaxy smartphone
By Jim Ford · #105378 ·