Re: 89441 Alignment/Test Disk ?
I doubt that Keysight would answer any questions regarding the 89441 as it is obsoleted
since a long time ago.
I have not seen FW 09.06 available or it is an unofficial/special version.
I also found
By
Ulf Kylenfall
·
#149939
·
|
Re: E4402 help please VCO Unlock.
This is the part I used to fix mine. Cheaper and easier to get. I think
Digi-Key or Mouser.
Sam Reaves
ARS W3OHM
Owner / Moderator of:
LeCroy Owners Group on 开云体育
Sencore Owners Group on
By
Sam Reaves
·
#149938
·
|
Re: E4402 help please VCO Unlock.
Excellent, really appreciate the help , i'll take the case back off and take a high res picture of the area around U54, I couldn't see it on the posted screenshot though? it was clearly visible on the
By
Tony G1HMO
·
#149937
·
|
Re: What VNA test port cables to get?
One cable brand I've had trouble with is MegaPhase.? According to my wireless lab manager friend, there is a manufacturing defect that causes suckouts.? They don't show up until you have used the
By
Jim Ford
·
#149936
·
|
Re: always-on AC power
The 2A fast-blow fuse offers sensible protection
You are perhaps talking about something OTHER than the 5316A. The 5316A has no fuse on the transformer secondary, and the primary-side mains fuse is
By
saipan59 (Pete)
·
#149935
·
|
Re: What VNA test port cables to get?
Be aware these are a consumable item. They gradually degrade in stability until they reach a point at which they are effectively useless. I keep meaning to put some information about how to check
By
Dr. David Kirkby, Kirkby Microwave Ltd
·
#149934
·
|
Locked
Re: Looking for an HP 85060 E-cal controller
On Thu, 6 Feb 2025, Bruce wrote:
I have a 85060C in a very good, almost pristine cosmetical condition. It
powers up and tries to boot from its floppy drive but that is as much as I
can tell about its
By
Sergey Kubushyn
·
#149933
·
|
Locked
Looking for an HP 85060 E-cal controller
Looking to acquire an HP 85060 E-cl controller - any condition as long as it is properly described.? Contact me off line if you have anything interesting.
I live in Western WA, US.
Chers!
Bruce
By
Bruce
·
#149932
·
|
Re: always-on AC power
FWIW, I have power-cycled my 5316A many times with the switch "on" - I have it plugged into a switched AC outlet.
Mine does not have GPIB, and the time base is the default simple crystal.
Is it
By
saipan59 (Pete)
·
#149931
·
|
Re: 89441 Alignment/Test Disk ?
Hi,
I have an 89441A (89410A/89431A) and an 89430A.
I have several original manuals and some diskettes, but unfortunately not
the adjustment diskette 89431-19403.
The rs-232 cable connecting
By
Yves Tardif
·
#149930
·
|
Re: always-on AC power
I recommend that you take care to switch these instruments off when you have to remove mains power. My 5316A counter blew the internal fuse on the transformer secondary, which failed by vaporising the
By
Julian_Hardstone
·
#149929
·
|
Re: E5071C Port Coupler Issue?
Just to clarify, the insertion loss test is as the image below, from the synth port (SMA) to the regular front panel VNA test port connector (N)
By
jmr
·
#149928
·
|
Re: E5071C Port Coupler Issue?
Can you also measure the VSWR of each RF port (synth input port and the external VNA test port) when you do the insertion loss measurement? This may indicate where a possible problem lies.
It ought
By
jmr
·
#149927
·
|
Re: E5071C Port Coupler Issue?
Googling for E5071C repairs, there's some images here in a youtube repair video.
TSP #174 - Teardown, Repair & Analysis of an Agilent 4.5GHz E5071C ENA Vector Network Analyzer (
By
jmr
·
#149926
·
|
Re: E5071C Port Coupler Issue?
Interesting circuit! Looking at the good one first...
SMD resistors marked 61X are = 42.2 ohm and 65X = 46.4 ohm.
So my guess it that this is a kind of sense resistor bridge (absorbed into a pi
By
jmr
·
#149925
·
|
Re: 89441 Alignment/Test Disk ?
My thought is that the analyzer may be configured as a standalone 89410A and so is not looking for the external RF unit when you connect it.
This may also be the cause of your missing menus
By
Peter Brown
·
#149924
·
|
Re: E5071C Port Coupler Issue?
Update:
Removed the board to see if there was anything on the back and... nix.? Solid ground plane.
I also checked out the working unit and noted that the components on the bottom of the array were
By
Scott Vance
·
#149923
·
|
Replacement for the 141T neon tube reference 1940-0013
Guys,
I have shipped all the BE19400013B neon tube replacement modules for the
HP141T I have in stock. I have more parts and boards. As soon as I can
build and test some more units, I will
By
Michael Bafaro
·
#149922
·
|
Re: 89441 Alignment/Test Disk ?
I have not used one of these for a while - I thought that the mode on the 10 MHz base unit had to be switched manually between 89410A and 89441A through the service menus?
Do you see an error message
By
Peter Brown
·
#149921
·
|
FS: HP/Agilent 16702B logic analyzer
For Sale: 16702B Logic Analyzer
Display is functional bug has an issue - LCD has lines in the display that diminish after warming up. External VGA does not have that issue.
OPT 003 - 256 Mb RAM
OPT
By
NickyDoes
·
#149920
·
|