开云体育

ctrl + shift + ? for shortcuts
© 2025 开云体育
Repair HP3455A Multimeter
Hi, my 3325A is driving me crazy and no one could help yet, so I went over to a faulty HP3455A multimeter. It worked many years, but now it comes up with a OL display in DC. Cal test stops at
By Ralf Rosche · #151076 ·
Re: DIY: Repair of HP Step Attenuators
Here’s a YouTube video I found when I was repairing the attenuator in my 8902a. ?I followed his procedure and used the same watch repair o ring kit from Amazon (under $10 shipped).
By pac1085 · #151075 ·
HP CRT "rough" equivalence info needed
I got a HP8754A recently that mostly works but seems low on available trace intensity. Maybe it just needs some internal adjustment, or maybe the CRT is weak on emission. Now that I have this one,
By ed breya · #151074 ·
Re: CAL TIMEBASE 8591A
I found the problem, I simply made a mistake in the procedure. Everything is correct now. Yves
By Yves Tardif · #151073 ·
Re: DIY: Repair of HP Step Attenuators
Hi Don I searched as you described but didn't find your repair procedure. What is the message number please? Thanks Jason NJ7K
By fishtronics · #151072 ·
Re: 89605 / 89606 cal file
It seems the cal files have one gain parameter starting at 50MHz in 50MHz steps. The smallest of the three files must’ve been intended for a system with an E2730 and the others for systems with
By Ed Marciniak · #151071 ·
CAL TIMEBASE 8591A
Hi, I have two 8591As to check, and for both, I'd like to correct the timebase. I followed the procedure in the service manual; it worked, and I correctly executed the CAL STORE command when
By Yves Tardif · #151070 ·
Re: DIY: Repair of HP Step Attenuators
Thanks Just ordered 50 ? orings.? I have 2 Attenuators to repair.
By ROBERT LEVY · #151069 ·
Re: DIY: Repair of HP Step Attenuators
How many of what size/type would be needed to replace all the ones used in an 8566B stack? I'd be interested in buying a set to refurbish my spec an when it becomes
By Dave Daniel · #151068 ·
Re: DIY: Repair of HP Step Attenuators
Is anyone going to be buying 0-rings now? I'd like to do a group order since I now have an old, mostly working 8566B SA. These can be mailed around in letters for $.73...
By John Griessen · #151067 ·
HP 435B, Zero Carryover Test - How noisy should it be?
Hey All, I noticed what looked to be a zero issue with my 435B, so I thought I'd run it through the zero carryover test (OSM 4-6) to see what was going on. While the meter meets all the ranges given
By Tony Goodhew · #151066 ·
Re: DIY: Repair of HP Step Attenuators
I used 0.5mm CS X 1.0mm ID 70 durometer from theoringstore.com on an attenuator from my HP
By Sam Reaves · #151065 ·
Re: DIY: Repair of HP Step Attenuators
I have done over a hundred repairs of the 3332xA/B style attenuators. In the file listing in this site you will find a detailed step by step procedure with HPAK part nos. and Apple Rubber part nos.
By Don Bitters · #151064 ·
Re: DIY: Repair of HP Step Attenuators
The original o-rings were 1.0mm ID,2.2mm OD for a cross section of 0.6mm. None of the ones listed are exactly the correct size.
By Ed Marciniak · #151063 ·
Re: DIY: Repair of HP Step Attenuators
There's good older threads that go into some detail on this. I repaired one set a few years ago, posting here, and I recall the "surgery" aspect of the project. Mechanically, maybe one of the more
By Radu Bogdan Dicher · #151062 ·
Re: DIY: Repair of HP Step Attenuators
Which size should I select on that page? I have stuck attenuator sections in three different instruments, seems to be a common problem. Marko Cebokli
By Marko Cebokli · #151061 ·
Re: DIY: Repair of HP Step Attenuators
HiI just disassembled my hp8648 attenuator 33322-60011. I found the little rubber orings have broke. I need to know exactly how to disconnect the rods that the orings are on from the contact, without
By ROBERT LEVY · #151060 ·
Re: 89605 / 89606 cal file
Awesome and thank you! I bricked two E1439C digitizers from a bad firmware update and developed a procedure to fix those. I need to write a small program to reset the serial number. The A1 (main
By Ed Marciniak · #151059 ·
Re: HP3325A Trouble, Calibration FAIL 2 & 3
I changed Q219 and its complementary transistor. Now I can adjust the 7.5Volts on F3. The amp is now fixed, but the A-CAL Fail error for square and triangle is still present. The output is correct,
By Ralf Rosche · #151058 ·
Re: Intermittent REF UNLOCK error message on 8591A Spectrum Analyzer
Hi Anders Just go to the Files section i have been uploaded the Clip to 8590/91A Br Johnny
By JLJ · #151057 ·