开云体育

ctrl + shift + ? for shortcuts
© 2025 开云体育
Re: Intermittent REF UNLOCK error message on 8591A Spectrum Analyzer
Hi Anders Just go to the Files section i have been uploaded the Clip to 8590/91A Br Johnny
By JLJ · #151057 ·
Re: E4402B SA power problem - need DC socket connections
Hello, Have a look on the messages " Agilent *E4404B* Spectrum Analyzer With Dead *PSU* Repair (
By f1ghb · #151056 ·
Re: microscope / high magnification camera options for soldering, faults, and other examinations/inspections
I later suspected that 15x would be too much. I'll have a look for AO570/145. Thanks. Renaud
By Renaud F5ZR · #151055 ·
Re: How to I disassemble the HP/Agilent 8495B attenuator ?
sorry, mine has N-type connectors and looks exactly as this one https://accusrc.com/product-Agilent-Keysight-8495B-10605
By Freiwat · #151054 ·
Re: How to I disassemble the HP/Agilent 8495B attenuator ?
Thanks, I have the manual, mine has no screws on each end, for? a? picture please see https://se.rs-online.com/web/p/programmable-step-attenuators/2237205
By Freiwat · #151053 ·
Re: microscope / high magnification camera options for soldering, faults, and other examinations/inspections
Hi Steve, Thanks for that advice. Why are 145 eyepieces better please? It seems 147 are very common. Best regards, Renaud The 145 eyepiece is 10x. The 147 is 15x. The AO570 body has magnifications
By Steve Holland · #151052 ·
Re: How to I disassemble the HP/Agilent 8495B attenuator ?
Do you have a manual attenuator or the electronically switched one? There are 4 screws on each end of the attenuator, if you have SMA/3.5mm connectors, then remove the nuts on those connectors, the
By Don Bitters · #151051 ·
Re: FS 8660 plug in's
Forgot to attach, sorry, Gary WA2OMY From: Gary Hitchner <talgarth@...> Sent: Monday, March 24, 2025 3:43 PM To:
By Gary Hitchner · #151050 ·
Anyone have the 17090C Measurement Graphics Software for the 7090A Plotter
Hey All, I asked over on VinHPcom but apart for a pointer to the 17090B Software up on HPMuseum I didn't hear much back, so I thought I'd ask here as well. The C version of the software apparently
By Tony Goodhew · #151049 ·
Re: FS 8660 plug in's
Do you have pictures? Glenn -- ----------------------------------------------------------------------- Glenn Little ARRL Technical Specialist QCWA LM 28417 Amateur Callsign:
By Glenn Little · #151048 ·
FS 8660 plug in's
List, FS - Three plug in's used in 8660's. 2)- 86603A one has phase modulation capability. These are the high frequency units that go to 2600 MHz. 1)- 86633B, Modulation plug in. These
By Gary Hitchner · #151047 ·
Re: E4402B SA power problem - need DC socket connections
The other area of the PSU to look at is standby power and startup circuit. In the older 859x series there is a Schottky diode HSMS-2800 that goes leaky. It's function is to rectify the output from a
By Chris Weir · #151046 ·
Re: E4402B SA power problem - need DC socket connections
That's good news. Robert.
By Robert G8RPI · #151045 ·
Re: E4402B SA power problem - need DC socket connections
Well, well, well!? I put 12V on the DC connector and removed the mains and the SA fired up.? about 12V at the plug at 7A when running, from a commercial switchmode 13.8V psu. And it stayed fired up
By Jeremy Owen · #151044 ·
Re: 89605 / 89606 cal file
Hi Ed, Here are 3 text files from separate serial number 89605A/B units.? I don't think I have any 606 cal files.? Would be happy to collaborate to set up a test procedure to build cal files for
By Adam, Shalimar FL · #151043 ·
Intermittent REF UNLOCK error message on 8591A Spectrum Analyzer
I have just bought an 8591A that has an intermittent REF UNLOCK error message when I use the internal 10 MHz reference. The error comes every 0,5 to 10 second. The internal reference is 10,000 MHz and
By Anders L??v · #151042 ·
Re: E4402B SA power problem - need DC socket connections
Thanks Robert.? That ties up with about 16.5V to chassis measured on that top pin.? The bottom pin seems to float. I'll try it on DC and see if that makes any difference.? I think that there's some
By Jeremy Owen · #151041 ·
How to I disassemble the HP/Agilent 8495B attenuator ?
Hello, I need to disassemble ?the HP/Agilent 8495B ?attenuator, there are no visible screws, how can I do it if I need to service the attenuator ? Thanks in advance!
By Freiwat · #151040 ·
Re: microscope / high magnification camera options for soldering, faults, and other examinations/inspections
Hi Steve, Thanks for that advice. Why are 145 eyepieces better please? It seems 147 are very common. Best regards, Renaud
By Renaud F5ZR · #151039 ·
Re: microscope / high magnification camera options for soldering, faults, and other examinations/inspections
Inasmuch as this is a vintage equipment group you might consider an American Optical AO570 with the model 145 eyepieces for about $100 on eBay. I can attest to the superb optics, with magnification
By Steve Holland · #151038 ·