Re: HP34702A Ohms converter issues
I hope so.? ;-) In this setup (Q12 / CR13 out of circuit), the gates of Q15 are offset just a bit - A: -5.0V B: -5.8V? The drains are of course then far apart - A: 2.8V B: 8.46V? and the sources
By gregdunn@... · #94322 ·
Re: HP34702A Ohms converter issues
The problem is obviously revolving around the op-amp which is made up of Q15 and U2. The two gates of Q15 should be balanced at all times, otherwise the output of U2 would be stuck at the +12 or the
By Joel Setton · #94321 ·
Re: VNA advice wanted
Pat, Thanks. I was pretty sure David had the A/B designations swapped, but confirmation greatly appreciated. And it would be very easy for someone reading the thread later to get things backwards with
By Reginald Beardsley · #94320 ·
Re: Interested in helping start/run a Wiltron/Anritsu group on groups.io?
Might take a few hours to be approved and propagate. [email protected]> wrote:
By bownes · #94319 ·
Re: Interested in helping start/run a Wiltron/Anritsu group on groups.io?
*[email protected] <[email protected]> *has been created. Have fun! [email protected]> wrote:
By bownes · #94318 ·
Re: HP34702A Ohms converter issues
With the protection circuitry disabled, the 1K resistor reads 1.87K. U3's inputs are both at -6.4V.? Its output is at -7.4V. Q15 seems to depend on what I'm measuring; for this test, its inputs are
By gregdunn@... · #94317 ·
Interested in helping start/run a Wiltron/Anritsu group on groups.io?
Looking for person (one or more) interested in starting and running a Wiltron/Anritsu equipment group (possibly modeled after this HP-Agilent-Keysight-equipment group) on groups.io. If you are
By David Feldman · #94316 ·
Re: VNA advice wanted
I bought mine from Globaltest a few years ago, I think I paid about $40 for it. -Dave -- Dave McGuire, AK4HZ New Kensington, PA
By Dave McGuire · #94315 ·
Re: VNA advice wanted
Dave, In regard to the collet extractor for APC-7, where did you find one cheap? I've been looking high and low for an inexpensive collet extractor without success. -Eric
By Eric Garner · #94314 ·
Re: Test fixture for dielectric measurements
Hmmm... If solid, you could hold the material in air away from the electrodes Chris Trask N7ZWY / WDX3HLB Senior Member IEEE http://www.home.earthlink.net/~christrask/
By Chris Trask - N7ZWY/WDX3HLB <christrask@...> · #94313 ·
Re: HP8714ET Impedance
Just going through my old emails. You may find the attached useful. Don't even think about getting into the bridge to change the impedance. HP uses stripline on a saphire substrate (on the 85046A
By peter bunge · #94312 ·
Re: VNA advice wanted
With all this great talk about VNA's it reminds me¡­ I have an extra 8753ES, 50ohm, internal s-para, working and looking beautiful¡­ 3ghz unit. $3k and also an n8357a 6ghz pna for sale¡­ same
By Pete Manfre <pmanfre@...> · #94311 ·
Re: HP34702A Ohms converter issues
Hello Greg, I think we're getting close to the root cause... but I thhink I need some more data. First, temporarily unsolder the collector lead of Q12 and one lead of CR13, which will disable the
By Joel Setton · #94310 ·
Re: VNA advice wanted
Yes, my error. I got the 85046A and 85046B transposed! The A is the 50 ohm version with APC7 connectors. I doubt the RF source, A, B or R receiver N connectors would get damaged as they are rarely
By Dr. David Kirkby from Kirkby Microwave Ltd <drkirkby@...> · #94309 ·
Re: Unleveled E4436B
Hi Tom, At 4GHz, using my E4406 to measure its output from -90dBm to +13dBm, it appears to be within 0.5dBm of indicated. Above 13dBm the power doesn't increase and the UNLEVELED indicator comes on at
By Tim Tuck · #94308 ·
Re: HP34702A Ohms converter issues
I see that the output of U3 is at -7.4V and the input is at -6.4V, so off from the specified value by 0.1 at each end.? The gate voltages of Q15 are finally balanced, which they weren't before, but
By gregdunn@... · #94307 ·
Re: VNA advice wanted
Just to be certain things don't get crossed up, the 85046A is a 50 ohm, 300 kHz to 3 GHz test set, and uses APC 7 connectors to interface with the DUT.? The 85047A is similar, but goes up to 6 GHz.
By Cubdriver · #94306 ·
Re: Test fixture for dielectric measurements
That is similar to what I see at NPL. Keysight and an aftermarket Swiss? company make a surface probe that can The coaxial probes can be made oneself, but I personally find the mathematics too hard to
By Dr. David Kirkby from Kirkby Microwave Ltd <drkirkby@...> · #94305 ·
Re: VNA advice wanted
[email protected] wrote: To my knowledge all *50 ohm* 3 GHz test sets (85046B) have APC7 connectors, not N connectors. If the test set has N connectors then it is almost certainly an 85046A which is
By Dr. David Kirkby from Kirkby Microwave Ltd <drkirkby@...> · #94304 ·
Re: VNA advice wanted
Thanks. I've got a lead on an 8753B with a 50 ohm 3 GHz S parameter test set at a comfortable price. Gedas gave me a very succinct explanation of why an eternal test set is preferable from a hobbyist
By Reginald Beardsley · #94303 ·