开云体育

Date

Re: HP8640B -140/+10dBm instead of -130/+20dBm?

 

Hello Thomas. Thanks very much for those useful references. It seems I won't be needing external amplifiers for my simple receiver testing. Jack at Clifton Labs mentions using LPFs to avoid confusion caused by generator harmonics, but mine are -60dBc so I hope added filtering will not be necessary.

Thanks everyone. Best wishes, Alan G3XAQ
--

Alan Ibbetson
alan@...


Re: Problem with 3MHz Filter in IF Section 8552B

 

Make sure the sweep rates used are commensurate with the IFBW - the shape will appear greatly distorted if sweeping too fast.

Ed


Re: HP-182C repair

 

Dave,

I have an HP182C with a 8559 spectrum analyser plugin. It too has developed a fault that I think is EHT-related.? It was working ok, then the brightness gradually increased but, the brightness control had no effect. I turned it off, waited 30 minutes and turned it on. Nothing showed on the screen.

I haven't stared the fault-finding yet, I was about to post in this forum and you got there first. You're obviously ahead of me in the game, but I wonder what your symptoms were?

73

Chris G3WIE


Looking for maual and or schematic for HP5303A counter

 

开云体育

Hello Group,

?

Recently I got a HP 5300 + 5303A counter for free with the comment: ”It does not work, maybe you can do something useful with it”

?

I opened it (very easy with these modular counters) and was greeted with? beautiful ?classic HP tech!

Also the reason for not working: an IC was completely missing! (It is IC 6 which is de-soldered)

It is a simple counter, but so beautifully made that I would like to restore it.

?

So, may question is: Can someone help me with the manual or schematics of this 5303A counter?

Please note: this counter is part of the 5300 series, and if you look on the Agilent/Keysight 5303A page there is a link to a manual, but the linked manual is for the 5300A, which contains no info whatsoever about the 5303A….. (I already received several 5300A manual from helpful people…..)

?

Alternatively, if somebody owns a 5303A, could they look inside to see what type IC6 is? (or make a photo of that print area)

(I did some testing, and the missing IC is the only thing wrong with this counter)

?

I started a topic about this counter on a Dutch forum: ?

I am afraid the text is in Dutch, but the nice Photo’s of the counter and it’s problem are international ;-)

(clicking the photo’s will show bigger versions)

?

Thanks for helping me out in advance, and regards,

Gertjan Miedema.

-----------------------------------------------------------------------------------
Miedema Geluidstechniek:


Re: Problem with 3MHz Filter in IF Section 8552B

 

Can you post a picture of what you are seeing?

Vladan


Problem with 3MHz Filter in IF Section 8552B

 

Hello,
the 3MHz filter shows the strange characteristic of a critically couple bandpass filter instead of the usual shape. Adjusting coils L2, L4 and R1 gives no improvement. The mica caps in this filter sh?uld be rock stable?! Before I desolder all components for measurement: does anyone have come across this problem before and can give me a hint.

Regards
Ernst


Re: HP8640B -140/+10dBm instead of -130/+20dBm?

 

For IMD measurements, you can use any coupler that can couple 2 or more generators together with some isolation.?

6dB Hybrid combiners are quite popular and broadband. Same goes for resistive couplers.?
A lot of interesting information on the different couplers can be found here:?

My personal favorite is a Wilkinson splitter. I do most of the measurements on VHF and UHF and with so high levels that a ferrite based coupler wil generate excessive IMD.

The isolation between the generators is normaly not a problem, with the 8640's you will most of the time use 20 or 30dB attenutation on both generators, leading to better than 60dB isolation, even without any isolation in the coupler.?
When measuring on receivers the input signals will often be low, but if measuring IMD on a power amplifier chain, then you may end up needing some external amplifiers that goes after the generators.

A set of good step attenuators after the combiner helps in setting the step level, and determining if the IMD is generated in the coupler or after it.

The book "Experimental methods in RF design" by Hayward, Campbell and Larkin have a good treatment of most IMD measurement setups.

BR.
Thomas LA3PNA.
?

2015-10-22 14:49 GMT+02:00 Alan Ibbetson alan@... [hp_agilent_equipment] <hp_agilent_equipment@...>:

?

Ah, so *that's* how the +20dBm level is reached. I now see the moving
cursor. I thought it was just a scratch on the panel! Thanks, Frank.

So that just leaves my question about the setup needed for receiver IMD
measurements.

Best wishes, Alan G3XAQ




--

?Please? avoid sending? me? Word? or? PowerPoint? attachments.
?See? <>
PDF is an better alternative and there are always LaTeX!


Re: HP8640B -140/+10dBm instead of -130/+20dBm?

 

Ah, so *that's* how the +20dBm level is reached. I now see the moving cursor. I thought it was just a scratch on the panel! Thanks, Frank.

So that just leaves my question about the setup needed for receiver IMD measurements.

Best wishes, Alan G3XAQ


Re: HP8640B -140/+10dBm instead of -130/+20dBm?

 

Hello,

On Thu, Oct 22, 2015 at 12:30 PM, Alan Ibbetson alan@...
[hp_agilent_equipment] <hp_agilent_equipment@...> wrote:


I'm new to this group. I've diligently searched the archives but can
find no mention of early 8640B models seemingly intended to produce
+10dBm maximum output instead of the more usual +20dBm.

I have a pair of these 8640B signal generators, the early-model one
having been recently purchased so I can make IMD measurements on my
homebrew ham radio receivers. The other one I've had for some years has
the common fault with broken plastic gears and Paul K is helping me out
with this, hence in the meantime me putting the later purchase into
service and being "surprised" at the lower output.

The serial number is 1434U00265, so it's a 1974 model. The output range
switch has a scale marking at -140dBm and also one at +20dBm, but the
switch does not rotate beyond +10dBm. Does anyone know when HP increased
the power output? Does it mean this example doesn't contain the dreaded
monolithic output amplifier? Although the range/deviation switches use
proper wafers instead of microscopic spring fingers on plastic discs the
mechanically mad plastic gears are there so I suspect there will be
service work to do in the future.
I have both models, the output level range is basicly the same on
both, but the one with the amplitude dial with the -140 dBm mark can
go to +20 dBm too using the fine level knob. You'll notice that the
level full scale will change as you move the fine level knob and a
small mark under the large dBm marks moves to indicate the actual
output power.

HTH
Frank IZ8DWF


HP8640B -140/+10dBm instead of -130/+20dBm?

 

I'm new to this group. I've diligently searched the archives but can find no mention of early 8640B models seemingly intended to produce +10dBm maximum output instead of the more usual +20dBm.

I have a pair of these 8640B signal generators, the early-model one having been recently purchased so I can make IMD measurements on my homebrew ham radio receivers. The other one I've had for some years has the common fault with broken plastic gears and Paul K is helping me out with this, hence in the meantime me putting the later purchase into service and being "surprised" at the lower output.

The serial number is 1434U00265, so it's a 1974 model. The output range switch has a scale marking at -140dBm and also one at +20dBm, but the switch does not rotate beyond +10dBm. Does anyone know when HP increased the power output? Does it mean this example doesn't contain the dreaded monolithic output amplifier? Although the range/deviation switches use proper wafers instead of microscopic spring fingers on plastic discs the mechanically mad plastic gears are there so I suspect there will be service work to do in the future.

Is it OK to ask applications questions on here? I'm wondering how much reverse isolation I need between my pair of generators to make accurate IMD measurements? I'll be looking for spurs at the noise floor of the receiver, say -130dBm or lower, and for a decent receiver this will need -30dBm signals from each generator. I'm thinking about, say, a 10dB pad between my hybrid combiner and receiver, with maybe a 20dB pad between each generator and the hybrid. Is this enough to ensure there will be no IMD from the generators caused by the crosstalk between them? Or do I need to think about external amplifiers and yet more attenuators?

Thanks,

Alan G3XAQ


Spectrum analyzer choice between ESA and PSA

 

Hi all,

Given the choice, would you rather pick an E4407B that is pretty optioned up, or a basic E4445A?

I don't need any of the personalities the ESA has and so the only attractive points are the narrow RBW and dynamic range expansion. I also don't really need the 26GHz BE, half is enough. Also, the Ethernet port on the PSA makes it easier for me to get data out of the device.

I used to have an E4406A and from what I hear the gui and use is very similar to the ESA.

I recently acquired an E8357A PNA, which is why I tend towards the PSA but I woul very much like to hear orher's opinions.

Thanks,
Chris


HP Rack mount, 5060-0174

 

Looking for rack mount kits for some of my HP gear, 5316, 53132A, 33401A, 437B, and similar sized gear.

Google doesn't find anything for 5060-0174.

Any suggestions?

bob k6rtm


Re: 11730B Cable "Mod" Query

 

On 10/21/2015 05:55 PM, w1ksz@... [hp_agilent_equipment] wrote:
I have three HP Power Meters (435B, 436A & 438A), three sensors (8482A,
8482H & 8484A)
along with three sensor cables (11730B).

One of the Sensor Cables has one of the pins removed such that is now a
7 pin Cable.
I was wondering why one of the pins would be removed ?

I cannot locate a pin-out diagram of the connector so cannot tell which
function it once
served (If anyone has a diagram of the connector pin-out. I would
appreciate it.

TIA for any help solving this,
Hi Dick, FYI the pinouts of these connectors are in the service
manuals of several of the power meter models.

-Dave

--
Dave McGuire, AK4HZ
New Kensington, PA


11730B Cable "Mod" Query

 

I have three HP Power Meters (435B, 436A & 438A), three sensors (8482A, 8482H & 8484A)
along with three sensor cables (11730B).

One of the Sensor Cables has one of the pins removed such that is now a 7 pin Cable.
I was wondering why one of the pins would be removed ?

I cannot locate a pin-out diagram of the connector so cannot tell which function it once
served (If anyone has a diagram of the connector pin-out. I would appreciate it.

TIA for any help solving this,


Dick, W1KSZ


Re: 8481A Repair

 

I would have spent the 400-600 dollars for a calibrated 437 and power sensor if I had not got mine much cheaper. The point……………i thought the same thing, but then realized that this is one of those times when simply paying the piper is the way to go. It is a VERY rare occasion that this is true for me, but here it is almost fact.

Cheers,

David


Re: 8481A Repair

Tom Miller
 

开云体育

That is because the repair is very expensive. And hard to find.
?

----- Original Message -----
Sent: Wednesday, October 21, 2015 2:23 PM
Subject: [hp_agilent_equipment] 8481A Repair

?

Has anyone had any experience repairing an 8481A power sensor? Functional ones look to be quite expensive but "needs repair" items are much more reasonable.

Dale


8481A Repair

 

Has anyone had any experience repairing an 8481A power sensor? Functional ones look to be quite expensive but "needs repair" items are much more reasonable.

Dale


Re: HP 215A manual

 

开云体育

Greetings,

?lists the HP-215A manual with schematics for $12.50.

This will be a very high quality scan. ?I have purchased a number of scans from them with total satisfaction.

Regards,

Ken


On 21Oct, 2015, at 12:55 AM, rbforum@... [hp_agilent_equipment] <hp_agilent_equipment@...> wrote:

Hi Dan and George,


I have a non-working 215A on my desk and besides of lacking enough old oval connector power cords I am in need of a manual.
So if you already have a scanned manual I would be glad to receive a copy of it.
Any help would be appreciated, even schematics only would help.

Thanks and regards,

Rudi



Re: HP 215A manual

 

开云体育

Rudi et all

If you dont find a free copy , we have an excellent scanned copy for download at a fair price



Cheers
Dave
ArtekManuals.com

On 10/21/2015 2:55 AM, rbforum@... [hp_agilent_equipment] wrote:
?

Hi Dan and George,


I have a non-working 215A on my desk and besides of lacking enough old oval connector power cords I am in need of a manual.
So if you already have a scanned manual I would be glad to receive a copy of it.
Any help would be appreciated, even schematics only would help.

Thanks and regards,

Rudi


-- 
Dave
Manuals@...



Re: HP-182C repair

 

HI My hp182c has a High voltage fault , The inverter transformer primary shows very Low resistace and the seconaries are Open circuit it will not oscillate

Can I get a high voltage inverter transformer antwhere ?


Dave? GM3WIL