Thanks, but been there, done that.? That info is also in the Assembly Level Repair Guide.?
I ordered the CLIP this morning, but I can't download it at work or on my phone.
That being said, the Hardware Failure 14 *IS* a P/S code; I found that info in the firmware manual.? I'll have to investigate further, but I'm going to change the damn battery first so it doesn't take 5 minutes to power it on.
Mike
WM4B
toggle quoted message
Show quoted text
On August 28, 2019 at 11:36 AM Siggi <siggi@...> wrote:
On Wed, Aug 28, 2019 at 7:04 AM Mike Besemer <
mwbesemer@...> wrote:
Thank you, Alwyn.? The backup battery has also entered my mind.? Where is it located?
The battery is a 3 cell NiCad on the processor board. If the battery's bad, the generator will self-calibrate on every power on. It's not the cause of the self-test failures.
I have the Artekmedia CLIP and service manual. On page 2c-63 of the service manual it lists the power supply error codes and the limits for the power supplies.