Re: 8970B Problem


kevin kearns
 

?
Gain reads -6.10db!

----- Original Message -----
Sent: Saturday, February 23, 2019 8:11 AM
Subject: Re: [HP-Agilent-Keysight-equipment] 8970B Problem

Hi All,?
?
?Does anyone else notice that it takes longer?for the NF reading to stabilize?compared with?the gain figure on the 8970B? When you calibrate the machine after a few minutes after switch on, gain reads zero and is stable. NF will read zero straight after cal?but drifts a bit, so it takes 10 mins warm up or so for the reading to?stay at zero after cal.
?
Checking the machine with a precision 6.0 dB attenuator, Gain reads 6.01dB and NF reads 6.10 dB when all warmed up.

Kevin
----- Original Message -----
Sent: Friday, February 22, 2019 11:26 PM
Subject: Re: [HP-Agilent-Keysight-equipment] 8970B Problem

Dear All,

Many thanks for your advice, especially to ¡°nj902¡± who pointed me to the explanation.

I¡¯ve now verified what happens as gain/NF of the measured object is changed and it is nice to find that this is not a fault.

Regards,

Alwyn

From:?nj902
Date: Thu, 21 Feb 2019 21:22:23 PST?

Apparently, this is normal.

Look up Agilent (Keysight) Application Note 1484? "Non-Zero Noise Figure After Calibration"

_____________________________________________________