You may, however, find in the cord either a white wire, black wire, and a green wire; or a blue wire a brown wire and a green wire.
If the former, white is neutral, black is live, and green is ground. If the latter, blue is live, brown is neutral, and green is ground.
toggle quoted message
Show quoted text
On Nov 2, 2018, at 19:35, Peter Gottlieb <
hpnpilot@...> wrote:
Oh, and it¡¯s wired as 240 center tapped, so the hot leads are interchangeable and there should be a green earth lead. No neutral in this case, with this 6-15 plug, here in the US power system.?
On Nov 2, 2018, at 7:23 PM, Dr. David Kirkby from Kirkby Microwave Ltd <
drkirkby@...> wrote:
Sorry, this is a bit off-topic for the HP/Agilent list, but I'm hoping someone, (probably from the USA), can help me.
I bought a Gamma high-voltage regulated power supply (0-15 kV, 0-12 mA).
from eBay.
The UK seller said it is 120 V, so needs a transformer to run from the UK mains, The mains voltage here is supposed to be 230 -6%/+10%. But I have my suspicions that the UK seller may be wrong, as the US manufacturer sent me a manual on the power supply, which says its 220 VAC. But they wrote it is a 20-year old custom design. (It's entirely possible someone has changed things.)
Am I right in thinking the plug attached to this is not designed for 120 V use, but instead for 240 V? When I look at it, with the earth pin at the top, the live/neutral are horizontal. The live/neutal pins are approximate 1/4" across, with approximately 1/2" between them, so there is approximately 1" between the two furthest edges of the pins.
I know one can always start at a low voltage, and I have a variac. But I'm also aware in some cases that using too low a voltage can damage equipment. So if this thing is supposed to take 220 V, I'd rather not start with 120 V.
The manual has no tolerance for the mains voltage - it just says 220 VAC. My own voltage is typically 235 V.
Some time ago the voltage was logged here for a period of 4 days and peaked at 255 VAC (RMS value), but the transformer taps have since been changed, and the voltage dropped 5%, so I think it might peak at 243 VAC RMS - and that would only be very occasionally.
One other thing. When I cut this mains plug off, what colour cables would I expect to see, and what would be live and neutral?
Here in the UK, neutral is very close in voltage to the earth wire - ideally, it is zero. Would there be an equivalent wire colour on this mains lead?
This thing can produce 15 kV at 12 mA, which I would guess would give someone a nasty burn at the very least, and could even be fatal. But my biggest concern at the minute is how to get power to this.?
Dave
--
Dr David Kirkby Ph.D C.Eng MIET
Kirkby Microwave Ltd
Registered office: Stokes Hall Lodge, Burnham Rd, Althorne, CHELMSFORD, Essex, CM3 6DT, United Kingdom.
Registered in England and Wales as company number 08914892
Tel 01621-680100 / +44 1621-680100