I will have to get back on what voltages I measured, however at the time of the incident which was some time ago, I recall from memory the fan came on when I pushed the power button but all the LED's on A17 were all out and there was no display on the front. This is after I realigned the power connector. I figured I probably blew something in the regulator. I should have measured the voltages coming out of the regulator test points but I did measure the voltages out of the rectifier board and they were all fine and within tolerance
I am currently testing Q10 which is 1884-0065 PRX, which is shown as 2N3670 in the manual. I performed the same test on it as the 2N4101 with the bulb and touching the gate, it works. However my concern is the current draw when I touch the gate it rises to my set current limit of about 0.4 to 0.5 on the DC supply and when I release it goes down and the bulb glows.