Re: Restoring HP 1740A scope, strange waveforms seen, assistance please?


 

¿ªÔÆÌåÓý

The signal is being differentiated, as if it is being coupled through a small value capacitor.? I would first look at the attenuator switch contacts on those particular v/div settings for an open or high resistance.? In other words dirty switch contacts.

Dave

On 10/15/2017 7:51 AM, graham.l.n.butcher@... [hp_agilent_equipment] wrote:

Hi all, I've been a member for a while but lurking in the background while working on the scope and looking out for any discussions about a 1740A taking place.


I now have the scope in a running condition after replacing 2 faulty transistors and now while testing the scope with the 1V test connection I discover that channel B works as expected but channel A works OK up to 0.2V on the volts/div switch, moving to 0.5V and upwards the trace goes odd.?


I was wondering if anyone could point me to the area that I need to be looking at on channel A to cure the problem. I have uploaded photos of? both channels with the settings on 0.5V and 1V so you can see the problem?


[url=][img][/img][/url][url=]B channel 1V dc[/url] by [url=]Graham Butcher[/url], on Flickr


[url=][img][/img][/url][url=]B channel 0.5V DC[/url] by [url=]Graham Butcher[/url], on Flickr


[url=][img][/img][/url][url=]A channel 1V DC[/url] by [url=]Graham Butcher[/url], on Flickr


[url=][img][/img][/url][url=]A channel 0.5V DC[/url] by [url=]Graham Butcher[/url], on Flickr


Thanks in advance for any help you can offer me.


Graham