mostly the Sampling Gate Diodes fail, and in very seldom cases
the following Fet, but normally the diodes protects the Fet
due to "fuse acting" before the Fet blows out.
You can substitute the Sampler Diodes with HP5082-2815
matched quad shottky diodes.
Self matching from fast shottkys is difficult but possible,
but they are matchet not only in forewardvoltage, as i remember right
also in capacitance as this affects the bridge balance, too.
The original Diodes have a Special HP-Number which is never found
again in any other document :-(, but the (now obselete, but here and
there still available, if in doubt, ask,i have some,) -2815 Quad works
well.
The HP 8405 Vector Voltmeter has the same Sampling Bridge and there
it also works well.
I have never understood why the probes are not field repairable-...
Hope that helps,
Jorgen
dj0ud
I have a HP 3406A Broadband Sampling Voltmeter I am repairing.
According to the test procedure in the manual , the probe assembly is
faulty and should be replaced as it is not field repairable.
I doubt if these are available from HP and if they were one could not
afford the price they would ask.
Does anyone have any experience with fixing these probes?