Well, I have found an important thing I had missed... the generator HAS output at some amplitude settings. I have then done the test on Service Sheet 8, page 8 of the Service Manual, to activate the attenuators, and when any of the AT2 (the 60dB unit) settings is selected, even the 0dB setting, signal output vanishes. All other settings, which only activate AT3 (the 80dB unit), work fine.
So it seems that the 60dB unit could be faulty. It is labelled 5061-4822. I wonder if there could be any parts unit around with that part in working condition. I would be very interested in getting one.