Re: S11 and S22 comparisons on 8753


David Kirkby
 

On 23 December 2012 15:11, Peter Bunge <bunge@...> wrote:
p.s. the cal kit discussion was very informative and timely. I still don't understand the comment (that was later corrected) about the open usually missing a part. My friend's Type N cal kit does not have extra parts for the open.
The cal kit the 8753 is supposed to work with is the 85032B - see the
8753 manual. The 85032B needs the extender, as I showed.



So unless your friend has a different cal kit, and has entered the
data as a user calibration, he is not getting the best accuracy. If he
has an 85032B, without the extender for the female N, all the data
from that device will be wrong.

Dave