开云体育

ctrl + shift + ? for shortcuts
© 2025 开云体育

Re: 3456A Voltmeter


 

Normal behaviour. They have VERY high input impedance on the low voltage ranges, so any stray charge can cause a reading.


Dave

-----Original Message-----
From: hp_agilent_equipment@... [mailto:hp_agilent_equipment@...] On Behalf Of Michael Hong
Sent: 20 December 2012 23:23
To: hp_agilent_equipment@...
Subject: [hp_agilent_equipment] 3456A Voltmeter

Hi everyone,
I just bought two 3456As from the same dealer off eBay.
Both of them have the same problem. When they are turned on, the number drift from low mV number and goes up to -8.nnnnn for one meter and the other meter to -6.nnnn. And then float around there. When I short the leads at Ohm or Volt setting, they show the near zero and when disconnected, go to their the number again.

I expect they stay at very low mV or zero without any input. What seems be the problem?
Thank you.


[Non-text portions of this message have been removed]



------------------------------------

Yahoo! Groups Links

Join [email protected] to automatically receive all group messages.