Are the Cal signal at w9 ok-about -25 ?—if so I guess you have the same fault we have seen in the MXA series-bad switch at Rf Front end..
Regards
Hardy
?
Fra:[email protected] <[email protected]> P? vegne af ipother6 via groups.io Sendt: 2. april 2025 16:46 Til:[email protected] Emne: [HP-Agilent-Keysight-equipment] RF Alignment Failure in Agilent CXA signal analyzer N9000A
?
I have Agilent CXA Signal Analyzer N9000A, I'm getting these errors, please help.
?
I referred manual and it suggested to check 50Mhz with -24dB, But when I check in my equipment its showing 50Mhz with -60dB. While I trace it out it was across H270(HMC270MS8G) where I was getting 50Mhz with -60dB but near FPGA it is exact 50Mhz with -24dB, so does it represent there is problem with H270 or any other if yes please help me, And any blue print of this will help a-lot.