Re: 34401A issue


 

Hi Peter,

If I am not mistaken there are 2 Omron relays on the board. If one
failed maybe the other one will fail soon so if you make an order and
have the unit open it is better to replace both.

Regards,
Razvan

On 09/03/2025 20:17, Peter Gottlieb via groups.io wrote:
I have tracked this down to K104 and confirmed by bench test the relay
has a bad side and won't make continuity to the DC high range
attenuator.? With the relay removed I can jumper pins 9 to 13 and get a
voltage reading on the higher range which confirms the rest of the meter
works.
This is an Omron G6AK-274P-STLT-US-5VDC which is commonly available.? I
will get one on order from Mouser.
Peter