HP3325A Trouble, Calibration FAIL 2 & 3


 

Hi,
I need some help for my HP 3325A, which worked many years perfect, but know has the common FAIL 2 & 3 Error. (PASS 1, FAIL 2 FAIL 3)
?
It has still sin, square and triangle output, the amplitude is ok, the range is ok (1-10Vpp) into a 50 Ohm load.
When swtichting from sin to square or triangle the a-cal error message comes up.
?
The voltages are good, 5.05, -15.16, +15.16 volts, no ripple, or minor. I changes some caps in the psu some time ago.
?
The integrator output looks good, the voltages are as expected on the testpoints LVL,OS1,OS2,AMPL,XDR. I interchanged already U20 & U24.
?
I have a revision C board, the realays on the att board are the flat black ones.
?
Alls tests were done grounding the ACD pin, 1khz and 10Vpp.
?
I found something strange in the amp section. placing the jumper to detach the amp from AMP IN and measuring the DC bias voltage at F3 left and right it shows -12.73 volts.
Should be +7.5 volts.? When toggeling the jumper and playing with the amp settings, the voltage sudden comes to a normal value and I could adjust the bias with the pot to 7.5 volts.
?
Should it be in any case 7.5 volts after power on and? when AMP in pin is disconnected and with any amp setting?
?
I checked already soldering and diodes, resistors, fuses and caps for shorts.?? And some transitors in the amp section.
?
I have the service manual, there is a huge section testing with a signature analyzer, but I don't have one.
?
So are there any hints, what to test / check next? Am I right in the amp section?
?
Ralf
?
?
?