Re: Agilent 34461A schenmatic


 

¿ªÔÆÌåÓý

I just discovered the partial schematic you attached.
Thank you! That will be of great assistance.


From: [email protected] <[email protected]> on behalf of Keith Harvey <keithh534@...>
Sent: Monday, October 7, 2024 12:05 PM
To: [email protected] <[email protected]>
Subject: Re: [HP-Agilent-Keysight-equipment] Agilent 34461A schenmatic
?
Hi Yves,

Thank you for the insight.

If you have the Agilent part numbers for the components I listed, that might prove helpful.

Regards,
Keith Harvey

From: [email protected] <[email protected]> on behalf of Yves Tardif <yves_tardif@...>
Sent: Monday, October 7, 2024 11:58 AM
To: [email protected] <[email protected]>
Subject: Re: [HP-Agilent-Keysight-equipment] Agilent 34461A schenmatic
?

Hi Keith,

?

I checked my 34410A and my 34465A, and unfortunately, I don't have a 34461A.

However, they all have the same function to perform, that is to say from +10V, provide a voltage of +3.3V, +1.2V and possibly +2.5V.

A switcher lowers a DC voltage from about +10V to +3.3V, the frequency of the switcher is about 1.3MHz for both instruments.

The 34461A has the same part numbers as the 34465A, there is a very good chance that the 34465A and 34461A have the same circuit for the section concerned.

?

I don't know if this can help you, but you have an idea of ??the damaged circuit.

I don't have the part numbers unfortunately, only the marking codes or the Keysight part numbers.

?

I repaired a 34461A a few months ago, the +3.3V switch was not damaged, but there was a short circuit on the +3.3V line caused by a faulty capacitor.

?

Waveform measured at output switcher connected to coil.

?

Yves