¿ªÔÆÌåÓý

Re: HP 8568A Spectrum Analyser not booting up. Both INSTR CHECK Leds on


 

I just realised something really important :
?
The service manual says "if the ADRS'D LED flashes without INSTR PRESET pressed it indicates A13 is functionning but cannot communicate with the controller processor"
?
The ADRS'D Led is indeed flashing on my unit... never realised it wasn't normal...
?
Btw, INSTR PRESET button is working just fine, i've tested it.
?
So that probably narrows it down to A15 right?
?
If i remove A13 (hpib board) it doesn't blink anymore but still got the same two instr check leds.
?
Also, i completly forgot to mention, if i flip the power off then on pretty fast i get random stuff on the screen, like lines, garbeled text-looking stuff and things like that. It always did that. They last for multiple seconds.
?
I just ordered a cheap usb logic analyser, i will do signature analysis checks.. in the meantime i'm searching for A15 and A14 boards online... people are selling them for more than double the price i paid for the whole unit (the 2 parts + the two cables + 4 volumes of manuals about it).
?
Seems cheaper to just buy a non-working RF part and try swapping the boards like you said...

Join [email protected] to automatically receive all group messages.