Re: 8566B Attenuator Service


 

¿ªÔÆÌåÓý

Okay, are you sure the internal switching contacts are working as intended?
It's been a few years since I was inside one but my memory is that there are two per solenoid?
Effectively a N/C one when the solenoid is in the 'de-selected' state and a N/O one that closes when it is selected.
They function in a break before make way where the power to the 'select' coil is routed through N/C contact and the power to the de-select coil through the N/O contact.
So, if the actuator rod is gunked up with gooey o-ring and not? moving freely over it's full travel I guess that might explain what you are seeing - which coil is remaining energised?

On 15/09/2024 12:20, Jinxie via groups.io wrote:

Okay. Well I have an 'intermediate update.' Cycling through 0 to 70dB I can see all solenoids switching seemingly normally - apart from the second from left in the photo. Commencing with all attenuators in the 'up'/off position, It flicks 'on' okay but then remains stuck there and won't release again. And clearly its coil is remaining energized as it can't be manually flicked off, either. Not sure if this could account for the problem I've had or whether there are other things amiss (such as the perished o-ring) also.
So, anyway, once that first 20dB attenuator is switched in, it cannot be switched out again and there's a minimum of 20dB attenuation permanently in-circuit. Could there be an issue with the control logic?