Hello Yves,
This second converter assembly is also used in the 8563E and also ESA
E4407B spectrum analyzers. In fact I think it is common for all
6.5/13.5/26.5GHz versions.
Mine had failed in the E4407B and it looks like it is related to an
incorrectly manufactured transformer inside the assembly.
I attached the Agilent Service Note related to this issue.
Part number affected is 5086-7958.
Thanks for the nice PDF with all explanations.
Regards,
Razvan
toggle quoted message
Show quoted text
On 13/09/2024 07:51, Yves Tardif via groups.io wrote:
Hi Adri,
A friend of mine had a faulty 8595E, the problem comes from the option
card 130.
But I took the opportunity to look at the second converter, so here are
some details on this module.
Regards
Yves