As Yves has said, there are no equivalent connectors on the 5351B. The 8350/83592 are quite old and the calibration procedure calls for equipment (5343A) of the same era.?
The more modern 5351B does not have the connectors you are looking for. That is the reason I keep a 5343A on my bench.?
On Sep 11, 2024, at 6:31?AM, Yves Tardif via groups.io <yves_tardif@...> wrote:
?
Hi,
?
On the 5351B there is no TRIGGER INPUT, only a function by pressing the TRIG button or via GPIB.
?
<image002.jpg>
<image006.png>
?
De?:[email protected] <[email protected]> De la part de i.brokos via groups.io ·¡²Ô±¹´Ç²â¨¦?: 11 septembre 2024 05:37 ??:[email protected] Objet?: Re: [HP-Agilent-Keysight-equipment] 5343A equivalency to 5351B
?
Hi Yves,
and thank you for the answer.
As you can see in the "Operating and service manual of Hp83592A plug-in"
in a Frequency Range and Accuracy Test page 4-3 the connection is
?
Hp8350B????????????????? Hp 5343A
CNTR TRIG??? <---> SWP INTFC A
STOP SWEEP <---> SWP INTFC B
?
What connection do I have to make if I use Hp5351B?