Re: 5343A equivalency to 5351B


 

¿ªÔÆÌåÓý

Hi,

?

I'm not sure I understand the question, but the 2 RF inputs on the front of a 5343A correspond to the two on the 5351B, as shown in the picture.

?

For the INTFC signal, it is an internal synchronization signal, between the 8350B and the plug-ins.

But to be able to make a measurement with a meter, you have to use the two BNC connectors 'CNTR TRIG' and 'STOP SWEEP'.

The problem is that the 5351B meter does not have an INTFC B input, but the 5343A speaker does.

So, if you have this meter, it would be possible to do it.

?

?

?

?

Yves

?

?

De?: [email protected] <[email protected]> De la part de i.brokos via groups.io
·¡²Ô±¹´Ç²â¨¦?: 10 septembre 2024 02:45
??: [email protected]
Objet?: [HP-Agilent-Keysight-equipment] 5343A equivalency to 5351B

?

Hello,

I'm trying to calibrate an HP8350B+Hp83592A plug in with an HP5351B

can anyone help me with the

?

2 terminals labelled SWP? INTFC?? A? &? B on HP5343A

?

which terminals are equivalent on HP5351B ? (if there is any)