I'm not sure I understand the question, but the 2 RF inputs on the front of a 5343A correspond to the two on the 5351B, as shown in the picture.
?
For the INTFC signal, it is an internal synchronization signal, between the 8350B and the plug-ins.
But to be able to make a measurement with a meter, you have to use the two BNC connectors 'CNTR TRIG' and 'STOP SWEEP'.
The problem is that the 5351B meter does not have an INTFC B input, but the 5343A speaker does.
So, if you have this meter, it would be possible to do it.
?
?
?
?
Yves
?
?
De?:[email protected] <[email protected]> De la part de i.brokos via groups.io ·¡²Ô±¹´Ç²â¨¦?: 10 septembre 2024 02:45 ??:[email protected] Objet?: [HP-Agilent-Keysight-equipment] 5343A equivalency to 5351B
?
Hello,
I'm trying to calibrate an HP8350B+Hp83592A plug in with an HP5351B
can anyone help me with the
?
2 terminals labelled SWP? INTFC?? A? &? B on HP5343A
?
which terminals are equivalent on HP5351B ? (if there is any)