¿ªÔÆÌåÓý

Re: More Investigation of the Malfunctioning Time Base RESET Lamp in my HP-1421A.


 

That's the action I see when using the LED circuit.? Yes, V105 illuminates as long as the RESET switch is pressed.? I don't have it in front of me, but I'm pretty sure mine is an 0803 prefix as well.

Thanks,
Barry - N4BUQ

My 1421A (serial 0803A03889), when I go to Single Sweep and press Reset switch S105, the Reset neon DS101 blinks and neon V105 illuminates as long as the switch is pressed.?

Jeremy?


On Sun, Apr 14, 2024 at 7:32?PM Jeremy Nichols via <jn6wfo=[email protected]> wrote:
Barry, what happens when you go to Single Sweep and press Reset switch S105??

My schematic (and yours) shows the Collector of Q109 at +48V but your measurement shows -0.33V??

Jeremy?


On Sun, Apr 14, 2024 at 7:20?PM n4buq via <n4buq=[email protected]> wrote:
IIRC, when I had the front panel off for cleaning, I inspected those and I _think_ they were very close.? Although I don't know why, they used 1% resistors there.? Of course, that doesn't prevent them from drifting but I do think they were quite close to spec.

Thanks,
Barry - N4BUQ

----- Original Message -----
> From: "Peter Gottlieb" <hpnpilot@...>
> To: "HP-Agilent-Keysight-equipment" <[email protected]>
> Sent: Sunday, April 14, 2024 9:07:32 PM
> Subject: Re: [HP-Agilent-Keysight-equipment] More Investigation of the Malfunctioning Time Base RESET Lamp in my
> HP-1421A.

> If you can lift Q109C then you could at least measure the sum of those two
> resistors.? But it would be nice to be able to measure the junction of them.
>
> On 4/14/2024 9:55 PM, n4buq via wrote:
>> I'll try grounding the collector of Q109.? Unfortunately, lifting one lead of
>> R136 isn't quick.? The front panel has to be removed to gain access to the
>> small assembly that holds DS101, R135, and R136.? It's too cramped to get to it
>> from the back side of the front of the chassis.
>>
>> Thanks,
>> Barry - N4BUQ
>>
>> ----- Original Message -----
>>> From: "Peter Gottlieb" <hpnpilot@...>
>>> To: "HP-Agilent-Keysight-equipment" <[email protected]>
>>> Sent: Sunday, April 14, 2024 8:10:33 PM
>>> Subject: Re: [HP-Agilent-Keysight-equipment] More Investigation of the
>>> Malfunctioning Time Base RESET Lamp in my
>>> HP-1421A.
>>> Let's go real simple.? You report the collector of Q109 is at essentially
>>> ground, right?
>>>
>>> So if you ground it, and ASSUMING R135 AND R136 HAVEN'T DRIFTED, the neon should
>>> see 86.9 volts, which should light it.
>>>
>>> Neons when new fire somewhere in the 65 volt range but when old can easily drift
>>> upwards, even higher than that voltage.? And resistors can and do drift out of
>>> tolerance.
>>>
>>> Check R135 and R136 values.? Then ground Q109 collector and measure the junction
>>> of R135 and R136 voltage.? If you see near 86.9 volts your neon is old and has
>>> too high a firing voltage.
>>>
>>> Quick and dirty, try lifting one lead of R136 and see if the neon lights then.
>>>
>>> Peter
>>>
>>>
>>> On 4/14/2024 8:59 PM, n4buq via wrote:
>>>> (New thread due to very incorrect subject line and opening paragraph in
>>>> previous thread.)
>>>>
>>>> I still have not determined why DS101 will not turn on in my HP-1421A
>>>> time-base.? I can ground Q109's base through a resistor and force the lamp to
>>>> turn on.? I can place a 1K resistor in parallel with R137 and also force the
>>>> lamp to turn on; however, nothing else works. I thought that the neon might
>>>> have aged to the point where there isn't sufficient striking voltage for it
>>>> but since it will turn on with these methods, that doesn't seem to be case(?).
>>>>
>>>> Setting the unit up as described in the manual for voltage tests, I get the
>>>> following:
>>>>
>>>> -100V F(B): -99.9V
>>>>
>>>> Q109
>>>> Emitter: -9.35V
>>>> Base: -8.78V
>>>> Collector: -0.33V
>>>>
>>>> Q110
>>>> Emitter: -8.01V
>>>> Base: -8.11V
>>>> Collector: -3.7V
>>>>
>>>> Given those values, it would appear (at least to me) that the lamp should turn
>>>> on but it doesn't.
>>>>
>>>> As a test, I disconnected R136/DS101 (the yellow (4) wire) from the board and
>>>> connected a battery, resistor, and LED between GND and the collector of Q109.
>>>> This works just as it should (i.e. the LED turns on when the sweep is
>>>> triggered, turns off at the end of the sweep, and repeats that).
>>>>
>>>> Is it possible that Q109 is not able to conduct enough current to turn DS101
>>>> on but enough to make the LED work?? That seems unlikely but, at this point, I
>>>> don't have any other explanation.
>>>>
>>>> Any help with this is appreciated.
>>>>
>>>> Thanks,
>>>> Barry - N4BUQ
>>>>
>>>>
>>>
>>>
>>>
>>
>>
>>
>>
>
>
>
>






Join [email protected] to automatically receive all group messages.