¿ªÔÆÌåÓý

Re: 5086-7337 *Yig Oscillator issue*


 

¿ªÔÆÌåÓý

Hi Ruben? I do know that the 10 meg to 20 Ghx is in 4 bands I was just checking you hadn¡¯t started looking at the fault wrongly ?

If I am fault finding on the Multiplier plug ins ??I start at the fundamental? yig frequency once that works then the others start to play

?

From: [email protected] [mailto:[email protected]] On Behalf Of RubenRubio
Sent: 12 March 2023 15:54
To: [email protected]
Subject: Re: [HP-Agilent-Keysight-equipment] 5086-7337 *Yig Oscillator issue*

?

Of course not

My mistake, those voltage values are present during mainframe command sweeps, in this case, from 10 meg to 20 gig. YO supose to do four sweeps at different freq spans to cover all bands.

Join [email protected] to automatically receive all group messages.