Re: 5086-7337 *Yig Oscillator issue*
Hi Ruben? I do know that the 10 meg to 20 Ghx is in 4 bands I was just checking you hadn¡¯t started looking at the fault wrongly ?
If I am fault finding on the Multiplier plug ins ??I start at the fundamental? yig frequency once that works then the others start to play
?
toggle quoted message
Show quoted text
From: [email protected] [mailto:[email protected] ] On Behalf Of RubenRubioSent: 12 March 2023 15:54To: [email protected] Subject: Re: [HP-Agilent-Keysight-equipment] 5086-7337 *Yig Oscillator issue*
?
Of course not My mistake, those voltage values are present during mainframe command sweeps, in this case, from 10 meg to 20 gig. YO supose to do four sweeps at different freq spans to cover all bands.
Are you sure you wish to delete this message from the message archives of
[email protected] ?
This cannot be undone.
Are you sure you wish to repost this message?