I had problems with a H75 option and I found a bad contact . I recommend to disassemble the unit carefully.?
Ing. Patricio A. Greco Taller Aeron¨¢utico de Reparaci¨®n 1B-349 Organizaci¨®n de Mantenimiento Aeron¨¢utico de la Defensa OMAD-001 Gral. Mart¨ªn Rodr¨ªguez 2159 San Miguel (1663) Buenos Aires T:?+5411-4455-2557 F:?+5411-4032-0072
On 27 Feb 2023, at 17:18, Paul Bicknell <admin@...> wrote:
?
Hi thank you for the return Emails
I will do a frequency sweep but we are talking down by several ?Dbm at 100 Mhz
I am 99.9 % convinced it is a problem with the series cap
Regards Paul
?
From:[email protected] [mailto:[email protected]] On Behalf Of Patricio A. Greco via groups.io Sent: 27 February 2023 21:27 To:[email protected] Subject: Re: [HP-Agilent-Keysight-equipment] HP 478A Thermistor head
?
How much goes down ? Did you sweep the sensor ? The input reflection coefficient is good tho check it.?
Ing. Patricio A. Greco Taller Aeron¨¢utico de Reparaci¨®n 1B-349 Organizaci¨®n de Mantenimiento Aeron¨¢utico de la Defensa OMAD-001 Gral. Mart¨ªn Rodr¨ªguez 2159 San Miguel (1663) Buenos Aires T:?+5411-4455-2557 F:?+5411-4032-0072
On 27 Feb 2023, at 11:24, Kenneth Harrison <bibpakak507@...> wrote:
?
Low end roll could be connector or bellows cap, guessing the later, but remote diagnostic is asking for trouble. Willing to help
?
On Mon, Feb 27, 2023, 7:43 AM Paul Bicknell <admin@...> wrote:
?
Hi? I have a HP 478A Thermistor head that works OK above 2 Ghz but reads low below 2 Ghz
I assume there is a problem with the internal series capacitor
Has anyone had a similar fault and is there any advice on a repair