¿ªÔÆÌåÓý

Re: HP-8753E Power Cycles at Start Up


 

The 5v supply MUST have a load on it or the other supplies will not be regulated?correctly (in the manual).
I am running it connected to the 8753B and an easy test is to 'scope the trip EN to the reset timer and add more load until it trips.
I can also jumper a resistor to force a trip as well.
It is worth learning about anyway.
Peter

On Mon, Aug 29, 2022 at 10:59 AM Rich Miller via <av8torrich=[email protected]> wrote:
Not to pile on - I have run the supply outside the case for over an hour. I see no ripple which would alarm me or make me suspect I have a bulk capacitance issue. I too am leaning towards overload trip point. I suppose I will have to figure out some way to load this supply down.?

Join [email protected] to automatically receive all group messages.