I realized a couple of hours after sending the initial post, that I was incorrect with associating Calibration with the Port Extensions.
Here is, I hope a better explanation of what I am trying to do.
First off the measurement frequency is 14 MHz.? So not too beyond the capability of the 8753C.
I want to place the measurement plane at the end of a 100 foot length of RG213.? So I place the SOL standards, one at a time at the
end of the cable.? Supposedly the 8753C measurement plane is now at the end of the cable.
Now I can take S11 measurements relative to the measurement plane at the end of the cable.? My question should have been
" Is there a limit in terms of how long a cable can be before this calibration technique fails."
This is probably related to the storage of the cal correction factors in the memory of the 8753C.? Is it via floating point numbers, or integer numbers?
How large or small can the calibration correction values be???
I seem to be able to obtain valid data using this technique, but wanted to ask on the Keysight forum with some of the folks who actually designed
the instrument.? I think Dr. Joel it still active there.?