On Sun, May 22, 2022 at 06:05 PM, John Quinn wrote:
Ozan, THANK YOU! The 100k resistor had failed open. I can¡¯t believe it. I'm slightly embarrassed, because I was sure I had already checked these. It is always a great idea to have a second set of eyes look at things. I¡¯ve been diagnosing this unit on and off for months. You have definitely helped keep an amazing piece of engineering out of the recycle pole for a 1 cent part.
Be well and have a wonderful week!
John Q
--------
Hi John,
Great news and congratulations for the repair. You already debugged it 99% of the way by identifying 1M-ohm from gate to ground makes a difference. That 100k resistor is not carrying any load so I wouldn't suspect it either if it wasn't for your observation with the 1M resistor.?
Ozan