¿ªÔÆÌåÓý

Re: 8516A cable question


Lothar baier
 

¿ªÔÆÌåÓý

There is no schematic for the 8516 only a few block diagrams , the switching isn¡¯t done by relays but rather with pin diodes inside the brick , the doubler control board is buttoned up on the Doubler , the only information to be gathered from the service manual is that there are 4 control lines that go to a PAL that acts as adress decoder , from there are 5 lines one goes to the doubler to switch it in and out the other 4 lines go to a block labeled filter control which outputs 5 lines that go to the actual doubler block so there is quite a bit of switching going on but there is no info on the switch points or details on the filters .

Having a crossover cable definatly might be a concern but he also mentioned that it used to work ¡°intermittent¡±? so that¡¯s also something to think about

?

?

Sent from for Windows

?

From: Bruce via groups.io
Sent: Wednesday, March 2, 2022 6:48 PM
To: [email protected]
Subject: Re: [HP-Agilent-Keysight-equipment] 8516A cable question

?

Sure but it is comparatively slow - Normal 8510C sweep is relatively?
slow, and the problem can be avoided by only weeping 50-20 and say?
25-40GHz.

I think the best first step is to look at the doubler switching module?
to see if the relay is actually switching the doubler in.? I don't?
seem to have a schematic diagram for the 8516, but hopefully, HP?
included a LED on the module?

In any case need to check the cable connections and determine whether?
the doubler is being energized - actually, maybe just watching the +?
supply to the doubler itself might work.

Cheers!

Bruce



Quoting Lothar baier <Lothar@...>:

> Disagree with you on this, dont forget that the synthesizer is swept?
> so if you do a 45MHz to 40GHz sweep the doubler is switched in and?
> out constantly so your DC now becomes a squarewave , also the?
> control signal not only switches the doubler but also switches in?
> and out filters to clean up the signal so there is some activity?
> going on those control lines !
> One way to find out would be to tack a scope probe on the control?
> lines and look at the shape of the control signals
>> On Mar 2, 2022, at 18:30, Bruce via groups.io?
>> <bruce@...> wrote:
>>
>> ?Missed that - the 20 ft cable seems NOT likely to be a problem -?
>> these ar DC control signals.? I recall some discussion that it is a?
>> straight through cable without cross over based on looking at 8341?
>> and 8516 schematic diagrams - I think that I'd focus on?
>> investigating the wiring of the cable.
>>
>> 1) Determine which conductors are used and be certain that the?
>> cable has those conductors 1-1 without cross over.
>>
>> Cheers!
>>
>> Bruce
>>
>> Quoting Lothar baier <Lothar@...>:
>>
>>> He mentioned earlier in this thread that its only 20-40GHz but he?
>>> also mentioned using a 20ft cable ? which could be the problem, I?
>>> am not sure what the control interface is , the part number of the?
>>> control cable says RS232 and there are 4 lines connected in the?
>>> testset BUT the doubler control uses a PAL as a decoder so I think?
>>> it¡¯s a parallel interface and having a 20ft cable could be a bit?
>>> much
>>>
>>> Sent from Mail<> for Windows
>>>
>>> From: Bruce via groups.io<mailto:bruce@...>
>>> Sent: Wednesday, March 2, 2022 6:21 PM
>>> To:?
>>> [email protected]<mailto:[email protected]>
>>> Subject: Re: [HP-Agilent-Keysight-equipment] 8516A cable question
>>>
>>> Does the IF error occur below 20 GHz ??? If not, the doubler is
>>> suspect - if so, look at the switching or the test set.
>>> Nexg step would be to try and look at the test set VTO.? It is
>>> difficult because they pipe it directly into the samplers.
>>>
>>> Cheers!
>>>
>>> Bruce
>>>
>>> Quoting ratn9ne@...:
>>>
>>>> Wow you guys are fast to help!
>>>> Here is some more info: error is only from 20-40GHz, so it has
>>>> something to do with the doubler I assume. Any span or frequency in
>>>> this range has this error. I have seen >20GHz work for a little
>>>> while but only randomly. My next idea is a shorter DB25 cable as
>>>> mine is about 20 feet long. Tried turning up the LO power coming in
>>>> on the sweeper with no luck. I can't find a replacement doubler on
>>>> ebay to test with. These must be very rare. Can they be serviced?
>>>> This display/sweeper/control box setup has been tested with an 8515
>>>> and no errors at all.
>>>>
>>>>
>>>>
>>>
>>>
>>>
>>>
>>>
>>>
>>>
>>>
>>>
>>>
>>
>>
>>
>>
>>
>>
>>
>>
>
>
>







?

Join [email protected] to automatically receive all group messages.