Hmmm have you unprotected the EEPROM ?? the E has the new flash based CPU board there is a little switch on the board that you have to toggle to remove the write protect.
I have fixed and upgraded well over 100 8753 and 8720 VNA when I was working at tucker and I don¡¯t recall any instance when the option codes did not work
I did not see any reference to the 8753E Series on Yahoo Groups. ?I did see some reference to the earlier versions to include those provided by Marcus.?
?
With regards to the Option Codes, I did get options code for both CPU¡¯s from Cesar. Both of them would not take these codes. This issue still puzzles me. I could understand a hardware fault for one not taking it, but two makes me thing
some of the units were locked down somehow.?
?
Rich
On Feb 21, 2022, at 3:55 PM, Lothar baier <Lothar@...> wrote:
?
Now obviously agilent released the option codes , there is a guy who provides you with the option codes , all you need to provide is the serial#
?
?
Sent from for Windows
?
From: Bruce via groups.io Sent: Monday, February 21, 2022 2:26 PM To: [email protected] Subject: Re: [HP-Agilent-Keysight-equipment] HP-8753E Options Activated
?
So where is the data you made public ??
Cheers
Bruce
Quoting Lothar baier <Lothar@...>:
> I figured this stuff out years ago on the 87xx as well as the?
> options on the 8360 and made it public, it cost me my job and I?
> nearly avoided a lawsuit from Agilent
?
> On Feb 21, 2022, at 13:26, Rich Miller via groups.io?
> <av8torrich@...> wrote:
>
> ? Hopefully someone can make use of this information. I found that?
> OPT 11 was involved by using 01 at that address. I also found that?
> using 01 addresses before the serial number would activate OPT10. I?
> did not get wise to the Fx Codes until Joe pointed them out to me.
>
> No worries about the flux, there is none left in the area. I used?
> lots of UltraSolve to clean the past flux I Utilize. This particular?
> flux is supposed to be safe if left, but I clean it anyways.
>
> On Feb 21, 2022, at 10:20 AM, Lothar@... wrote:
>
> ?The systematic is the same on pretty much all HP 8752/53/19/20/22?
> analyzers, you find the serial number and the 3 Fx codes are setting?
> the options!
> one word of caution though on the 8753E/ES CPU board, when removing?
> and resoldering the SMT eeprom
>? you need to clean up all flux residue throughly or you can get issues
>
>
>
>