Hi group ! I will start by saying that I really messed up here. I hope I will find help.
I noticed one day that my Analyzer was not reading the right amplitude in the 100Mhz range. It did read the correct amplitude at 300Mhz from the CAL output signal.
I decided to run the AMPTD CAL but it stopped with this error : AMPTD CAL : GAIN FAIL SRQ 110
Without fully reading the manuals, I then tried to run the CAL by centering to -37Hz. The AMPDT CAL then never finishes (after 12 hours of running).
I then restored the default CAL (I do have pictures of the CALS before I reloaded the default factory CALS).
?
Now, I can't run the CAL process without entering -37Hz because I get "CAL SIGNAL NOT FOUND" which is a documented problem during the calibration of the instrument, the workaround being to do the CAL by centering to -37Hz before.
?
Well, since centering to -37Hz is not working and now my CAL OUT at 300Mhz reads -50dbm instead of -20dmb, what can I do to fix this ?
?
If I could get back to the primary issue, which gave me the "AMPDT GAIN FAIL" (which is the DAC gain set to 255 but amplitude reading is not within specs), I could restart from where the real issue was.
Thanks.