¿ªÔÆÌåÓýRichard, The voltage at the R40/31/30 junction should be in the order of -180VDC with respect to the chassis. Note that this is the negative bias voltage generated across R40. If your voltage is +132V at R40/31/30 you have big troubles ¨C maybe check again to see if it is in fact negative. Next check the voltages at the cathodes of V2 and V4 these should be in the vicinity of -3.6VDC ¨C mine are -3.2 and -2.8 and the generator works OK. This is pin 3 on V2 and 4. Also check the grids of V2 and 4. Pin 2, ?and these should be around -19.5 or some such. Again mine are different at -16.5 and -16.7. Another check is to monitor the voltage at the junction of R40/31/30 whilst changing the Range switch from x100 to x 10K. there should be very little if any change. This checks to see if there are any issues with the two output transformers. One more thing ¨C on almost every 200CD that I have see C12 100uF NON POLARISED has gone bad. As we are only looking at the DV characteristics at present disconnect C12 9accross C30/31) and see if there are any changes to the DC bias voltages. ? Good luck Mark ? ? From: [email protected] <[email protected]>
On Behalf Of Richard Merifield via groups.io
Sent: Sunday, 7 November 2021 8:44 PM To: [email protected] Subject: Re: [HP-Agilent-Keysight-equipment] HP200CD oscillator ? I repaired L2 as best I could and it measured ok so thought I would give it a try. I suspect on 6AU6 may be part of my problem. I swapped them around and I now have a perfectly functional unit as far as I can tell, except R30 gets scolding
hot still. I have 4 more 6AU6 coming. |