Re: HP8562A Errors 304,335,317,333


 

¿ªÔÆÌåÓý

Would Like to say Good news but?

?

Managed to get the Analyser on the bench today for a quick check and looks as if 28Volts not there.

?

Also getting that smell of Burning rubber so worried could be causing more damage and need to isolate the source and start diagnosing supply.

?

Can the PSU work standalone, or is it load conscious?

?

Stu.

?

?

?

?

From: [email protected] <[email protected]> On Behalf Of Peter Hansen
Sent: 17 April 2020 09:51
To: [email protected]
Subject: Re: [HP-Agilent-Keysight-equipment] HP8562A Errors 304,335,317,333

?

hello Stu if you look on A2 on the side you see the LEDs for the PSU. All should be LIT. I am sure this is not a PSU issue.

If you work through the 300/600Mhz chain with an osilloscope you can check where the signal dissapears. I remember the 600Mhz is made in a doubler and if signal is too small you have not enough for doubling to 600Mhz.

best regards Peter

?


Fra: [email protected] <[email protected]> p? vegne af stur_mm via groups.io <stuart.ian.richardson@...>
Sendt: 17. april 2020 10:04
Til: [email protected] <[email protected]>
Emne: Re: [HP-Agilent-Keysight-equipment] HP8562A Errors 304,335,317,333

?

Peter,

?

Ok from schematic I can see the 600MHz Oscillator ¨¤ Buffer ¨¤ /2 = 300Mhz ¨¤ Offset Loop Buffer. Is that correct?

Is it the /2 device your talking about? The whole board seems to be a 600Mhz PLL with 10Mhz ref at source.

Correct me if I¡¯m wrong.

?

I still think something burn¡¯t out although th PSU looks clean.

Just prepping up before getting My hands dirty..

?

Stu.

?

From: [email protected] <[email protected]> On Behalf Of Peter Hansen
Sent: 16 April 2020 20:34
To: [email protected]
Subject: Re: [HP-Agilent-Keysight-equipment] HP8562A Errors 304,335,317,333

?

Hello Stu I must correct my self. The 8562A is before use og MMIC. area is still the same. It is just discrete amplifiers. Order a schematics from Artic media ant look for Radio Frequency module look in the page with the generation of 300/600Mhz.

Reason why you can?t measure anything is because the 600Mhz is used for third mixer.

best regards PEter

?


Fra: [email protected] <[email protected]> p? vegne af stur_mm via groups.io <stuart.ian.richardson@...>
Sendt: 16. april 2020 17:33
Til: [email protected] <[email protected]>
Emne: Re: [HP-Agilent-Keysight-equipment] HP8562A Errors 304,335,317,333

?

Peter,

?

Which board are we talking about ?

Still looking for the LED PSU indicators on the A2 card.

?

Presume you are referring RF board..?

Is this an off the shelf part or HP proprietary?

?

Thanks,

?

Stu.

?

From: [email protected] <[email protected]> On Behalf Of Peter Hansen
Sent: 16 April 2020 06:59
To: [email protected]
Subject: Re: [HP-Agilent-Keysight-equipment] HP8562A Errors 304,335,317,333

?

Hello Group a TAM is not needed in this case. It is simply a MMIC in the 300/600Mhz chain that is bad.

I believe Harke has a Hp85629A TAM for sale.

best regards Peter Hansen

?


Fra: [email protected] <[email protected]> p? vegne af Don Bitters via groups.io <donbitters@...>
Sendt: 16. april 2020 06:58
Til: [email protected] <[email protected]>
Emne: Re: [HP-Agilent-Keysight-equipment] HP8562A Errors 304,335,317,333

?

I suspect you might be able to rent the TAM from one of the ETE rental outfits for a month or so for not a horrible price.? I believe that Tam Hanna has one or two for sale, I may purchase one from him for my own use with my 8562A.
Don Bitters